Development of a Matrix-Matched Calibration Curve for Multi-Site Quantification of Neu5Gc-Bearing N-Glycans
Preprint
Abstract
N -glycolylneuraminic acid (Neu5Gc) has been repeatedly associated with human cancer, but reliable detection has remained elusive, generating controversy regarding its presence in human samples. To address this, matrix-matched calibration curves, which have been pioneered in proteomics and metabolomics for assessing changes in complex mixtures, were measured of released N -glycans at four orders of magnitude dynamic range in defined mixtures, systematically benchmarking Neu5Gc-containing N -glycan detection across multiple LC-MS platforms and sites. Orthogonally, the gold-standard analytical method, consisting of fluorescence detection of labelled monosaccharides separated by LC, was applied to the same samples, yielding absolute concentrations of Neu5Gc. LC-MS demonstrated an extended detection range of three or more orders of magnitude while retaining intact N -glycan measurement, improving assay specificity and enabling detection of the variety of Neu5Gc-bearing N -glycans. By combining orthogonal dimensions of evidence, including chromatographic separation, isotopic distribution matching, and composition-confirming MS/MS, LC-MS confidently resolved Neu5Gc signals from noise, even at low abundance. In comparison, DMB-LC-FLR was limited to two orders of magnitude dynamic range, insufficient for detection of Neu5Gc in commercially available pooled human sera. These findings strongly support that DMB-LC-FLR assay specificity and sensitivity are insufficient for Neu5Gc detection in human samples due to noise overwhelming the Neu5Gc signal. By establishing a reusable benchmarking framework for future glycomic studies, we aim to use LC-MS to improve the measurement of Neu5Gc in clinical samples.Keywords
Funding Information
- National Health and Medical Research Council (GNT2018947)
- National Health and Medical Research Council (GNT2026356)
This publication has 36 references indexed in Scilit:
- Elevation of serum N-glycolylneuraminic acid glycoconjugates in colorectal cancerBiochemical and Biophysical Research Communications, 2026
- An enhanced dual detection of DMB-labeled sialic acids using high-resolution accurate mass spectrometry and fluorescence detectionBBA Advances, 2025
- Extracellular vesicles display distinct glycosignatures in high-grade serous ovarian carcinomaBBA Advances, 2025
- Specificity and utility of SubB2M, a new N-glycolylneuraminic acid lectinBiochemical and Biophysical Research Communications, 2018
- Enhancing structural characterisation of glucuronidated O ‐linked glycans using negative mode ion trap higher energy collision‐induced dissociation mass spectrometryRapid Communications in Mass Spectrometry, 2017
- Using calibration approaches to compensate for remaining matrix effects in quantitative liquid chromatography/electrospray ionization multistage mass spectrometric analysis of phytoestrogens in aqueous environmental samplesRapid Communications in Mass Spectrometry, 2007
- Application of a sensitive fluorometric HPLC assay to determine the sialic acid content of infant formulasAnalytical and Bioanalytical Chemistry, 2007
- Fluorescent-Assisted Detection of Oligosialyl Units in GlycoconjugatesAnalytical Biochemistry, 1999
- An Introduction to Quadrupole Ion Trap Mass SpectrometryJournal of Mass Spectrometry, 1997
- High-pressure liquid chromatography of sialic acids on a pellicular resin anion-exchange column with pulsed amperometric detection: A comparison with six other systemsAnalytical Biochemistry, 1990