Safety of non-invasive brain stimulation in patients with implants: a computational risk assessment
Open Access
- 1 February 2025
- journal article
- Published by IOP Publishing in Journal of Neural Engineering
- Vol. 22 (1) , 016039
Abstract
Objective. Non-invasive brain stimulation (NIBS) methodologies, such as transcranial electric stimulation (tES) are increasingly employed for therapeutic, diagnostic, or research purposes. The concurrent presence of active/passive implants can pose safety risks, affect the NIBS delivery, or generate confounding signals. A systematic investigation is required to understand the interaction mechanisms, quantify exposure, assess risks, and establish guidance for NIBS applications. Approach. We used measurements, simplified generic, and detailed anatomical modeling to: (i) systematically analyze exposure conditions with passive and active implants, considering local field enhancement, exposure dosimetry, tissue heating and neuromodulation, capacitive lead current injection, low-impedance pathways between electrode contacts, and insulation damage; (ii) identify risk metrics and efficient prediction strategies; (iii) quantify these metrics in relevant exposure cases and (iv) identify worst case conditions. Various aspects including implant design, positioning, scar tissue formation, anisotropy, and frequency were investigated. Main results. At typical tES frequencies, local enhancement of dosimetric exposure quantities can reach up to one order of magnitude for deep brain stimulation (DBS) and stereoelectroencephalography implants (more for elongated passive implants), potentially resulting in unwanted neuromodulation that can confound results but is still 2–3 orders of magnitude lower than active DBS. Under worst-case conditions, capacitive current injection in the active implants’ lead can produce local exposures of similar magnitude as the passive field enhancement, while capacitive pathways between contacts are negligible. Above 10 kHz, applied current magnitudes increase, necessitating consideration of tissue heating. Furthermore, capacitive effects become more prominent, leading to current injection that can reach DBS-like levels. Adverse effects from abandoned/damaged leads in direct electrode vicinity cannot be excluded. Significance. Safety related concerns of tES application in the presence of implants are systematically identified and explored, resulting in specific and quantitative guidance and establishing basis for safety standards. Furthermore, several methods for reducing risks are suggested while acknowledging the limitations (see section 4.5).Keywords
This publication has 54 references indexed in Scilit:
- Limited output transcranial electrical stimulation 2023 (LOTES-2023): Updates on engineering principles, regulatory statutes, and industry standards for wellness, over-the-counter, or prescription devices with low riskBrain Stimulation, 2023
- Discussion on Spatial and Time Averaging Restrictions Within the Electromagnetic Exposure Safety Framework in the Frequency Range Above 6 GHz for Pulsed and Localized ExposuresBioelectromagnetics, 2019
- RF heating of deep brain stimulation implants in open‐bore vertical MRI systems: A simulation study with realistic device configurationsMagnetic Resonance in Medicine, 2019
- Modelling the electric double layer at electrode/electrolyte interfacesCurrent Opinion in Electrochemistry, 2019
- Low intensity transcranial electric stimulation: Safety, ethical, legal regulatory and application guidelinesClinical Neurophysiology, 2017
- Noninvasive Deep Brain Stimulation via Temporally Interfering Electric FieldsCell, 2017
- Safety of Transcranial Direct Current Stimulation: Evidence Based Update 2016Brain Stimulation, 2016
- Local tissue temperature increase of a generic implant compared to the basic restrictions defined in safety guidelinesBioelectromagnetics, 2011
- Pipeline embolization device (PED) for neurovascular reconstruction: initial experience in the treatment of 101 intracranial aneurysms and dissectionsNeuroradiology, 2011
- Selection of stimulus parameters for deep brain stimulationClinical Neurophysiology, 2004