NEW CONTRAST MECHANISM FOR SCANNING ELECTRON MICROSCOPE
Author:
- 15 February 1970
- journal article
- Published by AIP Publishing in Applied Physics Letters
- Vol. 16 (4) , 151-153
Abstract
With normal electron incidence, the resolution of the backscattered electron image in the scanning electron microscope (SEM) is approximately equal to the classical electron pentration depth. With oblique electron incidence, a significant number of plurally scattered electrons leave the specimen in an apparently specular direction after penetrating for a distance that is an order of magnitude smaller than this. Thus with 15-keV electrons incident onto Al at 45°, a significant number of backscattered electrons leave the specimen after penetrating to less than 500-a.u. depth. These electrons can be collected over an angle that is close to the plane of the specimen surface. Other electrons leave the specimen more nearly at right angles to the surface, and these have been scattered from a greater depth. The image in the SEM can change completely if the position of the collector is changed.Keywords
This publication has 6 references indexed in Scilit:
- Multiple scattering of 5 - 30 keV electrons in evaporated metal films III: Backscattering and absorptionBritish Journal of Applied Physics, 1965
- Multiple scattering of 5-30 keV electrons in evaporated metal films II: Range-energy relationsBritish Journal of Applied Physics, 1964
- Multiple scattering of 5-30 keV electrons in evaporated metal films: I. Total transmission and angular distributionBritish Journal of Applied Physics, 1964
- The plural scattering of 20 kev electronsBritish Journal of Applied Physics, 1964
- Wide-band detector for micro-microampere low-energy electron currentsJournal of Scientific Instruments, 1960
- New Method for Range Measurements of Low-Energy Electrons in SolidsJournal of Applied Physics, 1959