Evaluation of the PathoTec "Rapid I-D System".
Authors:
- D. J. BlazevicDBD. J. Blazevic
- P. C. SchreckenbergerPSP. C. Schreckenberger
- J. M. MatsenJMJ. M. Matsen
- 1 December 1973
- journal article
- Vol. 26 (6) , 886-9
Abstract
The 10 biochemical test strips included in the PathoTec Rapid I-D System were evaluated for accuracy as compared to standard tests and for efficacy in identification of 193 gram-negative bacilli. The test agreement was 100% for oxidase and phenylalanine deaminase, 99% for indole, nitrate, and Voges-Proskauer, 98% for malonate, 97% for lysine decarboxylase, 90% for urease, 84% for H(2)S, and 75% for esculin hydrolysis. Most of the commonly isolated Enterobacteriaceae were identified correctly within 4 h. Errors in identification of Proteus morganii and P. rettgeri occurred because of positive H(2)S tests on the PathoTec strips with these organisms.Keywords