Abstract
Matrix effects (MEs) can adversely affect quantification in pesticide residue analysis using GC. Analyte protectants (APs) can effectively interact with and mask active sites in the GC system, and are added individually or in combination to sample extracts and calibration solutions to minimize errors related to MEs. Unfortunately, APs cannot sufficiently compensate for MEs in all cases. Plant extracts, containing a broad range of natural compounds with AP properties, can also be used for this purpose. In this study, the applicability of cucumber extract as a natural AP mixture was investigated both alone and in combination with traditional APs. Extracts of two selected difficult matrices (onion and garlic) were prepared according to the citrate-buffered QuEChERS (quick, easy, cheap, effective, rugged, and safe) procedure. ME values of 40 representative GC-amenable pesticides were compared when calibrating against standards in pure solvent and in cucumber extract, with and without the addition of APs. Using a GC system with a contaminated inlet liner, the use of a cucumber-based calibration solution decreased MEs remarkably. The combination of APs with cucumber raw extract further decreased MEs, resulting in more than 85% of the tested pesticides showing ≤ 10% ME in onion and ≤ 20% ME in garlic. These results demonstrate that the preparation of calibration standards based on cucumber extracts (with or without the addition of APs) is a very useful and practical approach to compensate for MEs in pesticide residue analysis using QuEChERS and GC-MS/MS. The use of various internal standards is furthermore critically discussed.Keywords
Funding Information
- National Institute of Agricultural Sciences (NAS) of the Rural Development Administration (Project No. PJ012217)
This publication has 24 references indexed in Scilit:
- Compensation of matrix effects in gas chromatography–mass spectrometry analysis of pesticides using a combination of matrix matching and multiple isotopically labeled internal standardsJournal of Chromatography A, 2017
- Internal standards for use in the comprehensive analysis of polychlorinated aromatic hydrocarbons using gas chromatography combined with multiphoton ionization mass spectrometryJournal of Chromatography A, 2016
- Validation and use of a QuEChERS-based gas chromatographic–tandem mass spectrometric method for multiresidue pesticide analysis in blackcurrants including studies of matrix effects and estimation of measurement uncertaintyTalanta, 2014
- Variability of matrix effects in liquid and gas chromatography–mass spectrometry analysis of pesticide residues after QuEChERS sample preparation of different food cropsJournal of Chromatography A, 2012
- Assessment of sample cleanup and matrix effects in the pesticide residue analysis of foods using postcolumn infusion in liquid chromatography–tandem mass spectrometryJournal of Chromatography A, 2011
- Compensation for matrix effects in gas chromatography–tandem mass spectrometry using a single point standard additionJournal of Chromatography A, 2009
- Application of the standard addition approach for the quantification of urinary benzeneJournal of Chromatography B, 2005
- Matrix effects in (ultra)trace analysis of pesticide residues in food and biotic matricesJournal of Chromatography A, 2003
- Does further clean-up reduce the matrix enhancement effect in gas chromatographic analysis of pesticide residues in food?Journal of Chromatography A, 2000
- Explanation of the matrix-induced chromatographic response enhancement of organophosphorus pesticides during open tubular column gas chromatography with splitless or hot on-column injection and flame photometric detectionJournal of Chromatography A, 1993