An In Vitro Study of the Accuracy and Precision of Breathalyzer® Models 900, 900A, and 1000
Authors:
- 1 October 1985
- journal article
- research article
- Published by ASTM International in Journal of Forensic Sciences
- Vol. 30 (4) , 1058-1063
Abstract
Ninety Breathalyzer® instruments (Model 1000) and twenty instruments (Models 900, 900A) were studied using a protocol described by the Department of Transportation's “Standard for Devices to Measure Breath Alcohol.” Although the mean of each of three concentrations tested (0.05, 0.10, and 0.15 g/210 L) compared favorably in both series, the standard deviation was consistently higher for the Model 1000 instruments. The Model 1000 instruments also produced a significant number of test results which exceeded the normally expected scientific deviation.Keywords
This publication has 2 references indexed in Scilit:
- In Vitro Studies on a BreathalyzerJournal of Forensic Sciences, 1972