A search for optimal x-ray spectra in iodine contrast media mammography
Authors:
- Michael SandborgMSMichael Sandborg
- David R DanceDDDavid R Dance
- Martin YaffeMYMartin Yaffe
- Gudrun Alm CarlssonGCGudrun Alm Carlsson
- 22 June 2005
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 50 (13) , 3143-3152
Abstract
The aim of this work was to search for the optimal x-ray tube voltage and anode-filter combination in digital iodine contrast media mammography. In the optimization, two entities were of interest: the average glandular dose, AGD, and the signal-to-noise ratio, SNR, for detection of diluted iodine contrast medium. The optimum is defined as the technique maximizing the figure of merit, SNR2/AGD. A Monte Carlo computer program was used which simulates the transport of photons from the x-ray tube through the compression plate, breast, breast support plate, anti-scatter grid and image detector. It computes the AGD and the SNR of an iodine detail inside the compressed breast. The breast thickness was varied between 2 and 8 cm with 10-90% glandularity. The tube voltage was varied between 20 and 55 kV for each anode material (Rh, Mo and W) in combination with either 25 microm Rh or 0.05-0.5 mm Cu added filtration. The x-ray spectra were calculated with MCNP4C (Monte Carlo N-Particle Transport Code System, version 4C). A CsI scintillator was used as the image detector. The results for Rh/0.3 mmCu, Mo/0.3 mmCu and W/0.3 mmCu were similar. For all breast thicknesses, a maximum in the figure of merit was found at approximately 45 kV for the Rh/Cu, Mo/Cu and W/Cu combinations. The corresponding results for the Rh/Rh combination gave a figure of merit that was typically lower and more slowly varying with tube voltage. For a 4 cm breast at 45 kV, the SNR2/AGD was 3.5 times higher for the Rh/0.3 mmCu combination compared with the Rh/Rh combination. The difference is even larger for thicker breasts. The SNR2/AGD increases slowly with increasing Cu-filter thickness. We conclude that tube voltages between 41 and 55 kV and added Cu-filtration will result in significant dose advantage in digital iodine contrast media mammography compared to using the Rh/Rh anode/filter combination at 25-32 kV.Keywords
This publication has 17 references indexed in Scilit:
- Human observer detection experiments with mammograms and power‐law noiseMedical Physics, 2001
- Influence of anode/filter material and tube potential on contrast, signal-to-noise ratio and average absorbed dose in mammography: a Monte Carlo study.British Journal of Radiology, 2000
- Contrast and dose with Mo-Mo, Mo-Rh, and Rh-Rh target-filter combinations in mammography.Radiology, 1995
- Optimization of spectral shape in digital mammography: Dependence on anode material, breast thickness, and lesion typeMedical Physics, 1994
- Calculation of dose and contrast for two mammographic gridsPhysics in Medicine & Biology, 1992
- Influence of anode and filter material on image quality and glandular dose for screen-film mammographyPhysics in Medicine & Biology, 1991
- Computation of bremsstrahlung X-ray spectra and comparison with spectra measured with a Ge(Li) detectorPhysics in Medicine & Biology, 1979
- Absorbed Radiation Dose in MammographyRadiology, 1979
- Relativistic atomic form factors and photon coherent scattering cross sectionsJournal of Physical and Chemical Reference Data, 1979
- Atomic form factors, incoherent scattering functions, and photon scattering cross sectionsJournal of Physical and Chemical Reference Data, 1975