Is liquid heparin comparable to dry balanced heparin for blood gas sampling in intensive care unit?
Authors:
Open Access
- 1 January 2014
- journal article
- research article
- Published by Jaypee Brothers Medical Publishing in Indian Journal of Critical Care Medicine
- Vol. 18 (1) , 14-20
Abstract
Introduction: Blood gas (BG) analysis is required for management of critically ill patients in emergency and intensive care units. BG parameters can be affected by the type of heparin formulations used-liquid heparin (LH) or dry balanced heparin (DBH). This study was conducted to determine whether blood gas, electrolyte, and metabolite estimations performed by using DBH and LH are comparable. Materials and Methods: A prospective study was conducted at pediatric intensive care unit (PICU) of a tertiary care hospital. Paired venous samples were collected from 35 consecutive children in commercially prepared DBH syringes and custom-prepared LH syringes. Samples were immediately analyzed by blood gas analyzer and compared for pH, pCO 2 , pO 2 , HCO 3 - , Na + , K + , Cl - , and lactate. Paired comparisons were done and agreement was assessed by Bland-Altman difference plots. The 95% limits of absolute agreement (LOA) were compared with the specifications for total allowable error (TEa). Results: The P values were significant for all measured parameters, with the exception of pCO 2 and K +. Bland-Altman difference plots showed wide LOA for pCO 2 , pO 2 , HCO3 - , Na + , K + , and Cl - when compared against TEa. For pCO 2 , HCO3 - , Na + , K + , and Cl - , 40%, 23%, 77%, 34%, and 54% of samples were outside the TEa limits, respectively, with LH. Conclusion: Our study showed that there is poor agreement between LH and DBH for the BG parameters pCO2, pO2, HCO3 - , K + , Na + , and Cl - and, thus, are not comparable. But for pH and lactate, LH and DBH can be used interchangeably.Keywords
This publication has 22 references indexed in Scilit:
- A Comparative Study on Reliability of Point of Care Sodium and Potassium Estimation in a Pediatric Intensive Care UnitIndian Journal of Pediatrics, 2013
- Effects of Blood Clots on Measurements of pH and Blood Gases in Critical Care AnalyzersPoint of Care: The Journal of Near-Patient Testing & Technology, 2011
- Impact of blood collection devices on clinical chemistry assaysClinical Biochemistry, 2010
- Heparin release is insufficient in syringes with platelets as heparin sourceClinica Chimica Acta, 2008
- Liquid Heparin Anticoagulant Produces More Negative Bias in the Determination of Ionized Magnesium than Ionized CalciumYonsei Medical Journal, 2006
- Current databases on biological variation: pros, cons and progressScandinavian Journal of Clinical and Laboratory Investigation, 1999
- Comparison of the Effects of Dry and Liquid Heparin on Neonatal Arterial Blood GasesAmerican Journal of Perinatology, 1992
- Effect of heparin on arterial blood gasesAnnals of Emergency Medicine, 1985
- Too much heparin: possible source of error in blood gas analysis.BMJ, 1983
- Effect of Heparin or Saline Dilution of Blood on Pco2and pHUpsala Journal of Medical Sciences, 1975