Reconstruction of 4‐MV bremsstrahlung spectra from measured transmission data
Authors:
- Pin‐Hua HuangPHPin‐Hua Huang
- Kenneth R. KaseKKKenneth R. Kase
- Bengt E. BjärngardBBBengt E. Bjärngard
- 1 November 1983
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 10 (6) , 778-785
Abstract
Transmission data for 4‐MV bremsstrahlung beams have been measured with a combination of lead and aluminum attenuators. From these data, the original energy spectra have been reconstructed using an iterative least‐squares technique, previously evaluated by simulation studies. The spectra on the central axis for three similar 4‐MV linear accelerators indicated no significant differences. When studying the spectra at 5° and 9° off the central axis, that at 9° showed the expected increase of low‐energy photons. All these spectra showed a maximum photon energy of 4.5±0.2 MeV. When the magnetron power was reduced, the spectrum on the central axis shifted to lower energies and the maximum photon energy decreased to 3.5±0.2 MeV. The result of this experimental study confirms the conclusions from the previous simulation, that the numerical technique for analysis of transmission data can accurately represent 4‐MV bremsstrahlung spectra and detect differences in energy distribution with changes in machine tuning and position in the radiation field for a 4‐MV bremsstrahlung beam.Keywords
Funding Information
- National Cancer Institution (CA 28959)
This publication has 7 references indexed in Scilit:
- Spectral characterization of 4 MV Bremsstrahlung by attenuation analysisMedical Physics, 1981
- X‐ray spectra vs attenuation data: A theoretical analysisMedical Physics, 1981
- Current collection and ionic recombination in small cylindrical ionization chambers exposed to pulsed radiationBritish Journal of Radiology, 1980
- Off-axis beam quality change in linear accelerator x-ray beamsMedical Physics, 1980
- Photon Energy Spectra of a Heavily Filtered 300 kV Therapy UnitActa Radiologica: Oncology, 1980
- X-Ray Spectra from Accelerators in the range 2 to 6 MeVPhysics in Medicine & Biology, 1967
- Saturation Characteristics of Parallel-Plate Ionization ChambersPhysics in Medicine & Biology, 1964