Influence of Corneal Biomechanical Properties on Intraocular Pressure Differences Between an Air-Puff Tonometer and the Goldmann Applanation Tonometer
Authors:
- Marco LombardoMLMarco Lombardo
- Francesco OddoneFrancesco Oddone
- Sebastiano SerraoSSSebastiano Serrao
- Domenico Schiano LomorielloDomenico Schiano Lomoriello
- Pietro DucoliPDPietro Ducoli
- 1 January 2013
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Glaucoma
- Vol. 22 (5) , 416-421
Abstract
To estimate the influence of corneal properties on intraocular pressure (IOP) differences between an air-puff tonometer (NT530P; Nidek) and the Goldmann applanation tonometer (Haag-Streit). The influence of central corneal thickness (CCT), keratometry, and Ocular Response Analyzer (Reichert) measurements of corneal viscoelasticity [corneal hysteresis (CH) and corneal resistance factor (CRF)] on IOP differences between tonometers was evaluated. The CRF was calculated to be the best predictor of the differences in IOP readings between tonometers (r2=0.23; P<0.001), followed by CCT (r2=0.15; P=0.003) and CH (r2=0.14; P=0.003). Keratometry performed very poorly as lone predictor of IOP differences. In a multiple regression model, CRF, CH, and CCT together accounted for 25% (r2=0.25; P<0.01) of the variance in IOP reading differences between tonometers. Corneal resistance to applanation induced by either contact or noncontact tonometers was calculated to be the most determinant factor in influencing IOP differences between applanation tonometers.Keywords
This publication has 10 references indexed in Scilit:
- Ocular Response Analyzer and Goldmann Applanation TonometryJournal of Glaucoma, 2010
- On repeated corneal applanation with the Goldmann and two non‐contact tonometersClinical and Experimental Optometry, 2010
- Biomechanical properties of the cornea measured by the Ocular Response Analyzer and their association with intraocular pressure and the central corneal curvatureClinical and Experimental Optometry, 2009
- Evaluation of the Influence of Corneal Biomechanical Properties on Intraocular Pressure Measurements Using the Ocular Response AnalyzerJournal of Glaucoma, 2006
- Determining in vivo biomechanical properties of the cornea with an ocular response analyzerJournal of Cataract & Refractive Surgery, 2005
- Influence of corneal biomechanical properties on intraocular pressure measurementJournal of Cataract & Refractive Surgery, 2005
- Human Corneal Thickness and Its Impact on Intraocular Pressure MeasuresSurvey of Ophthalmology, 2000
- Comparison of the Performance of the Nidek NT-2000 Noncontact Tonometer with the Keeler Pulsair 2000 and the Goldmann Applanation TonometerOptometry and Vision Science, 1997
- Sources of error with use of Goldmann-type tonometersSurvey of Ophthalmology, 1993
- APPLANATION TONOMETRY AND CENTRAL CORNEAL THICKNESSActa Ophthalmologica, 1975