Coronary balloon catheter tip damage. A bench study of a clinical problem
Open Access
- 8 December 2017
- journal article
- research article
- Published by Wiley in Catheterization and Cardiovascular Interventions
- Vol. 92 (5) , 883-889
Abstract
To confirm clinically that coronary balloon catheter tips may be damaged during bifurcation treatment with side-branch access through the side of a stent. On the bench, we aimed to assess the susceptibility of different balloon designs to damage. We compared catheter tip widths. We tested whether balloon tip flaring can cause stent distortion. We had observed that balloon catheters that failed to cross to a side-branch frequently exhibited tip damage. We examined microscopically for damage 82 balloon tips after clinical side-branch access. In a bench study, the forces required to compress catheter tips 0.5 mm were compared to assess susceptibility to damage. We compared tip widths of balloons of different nominal inflation diameters. We examined stents after side-branch access for distortion. In 42 of 48 (88%) of balloon tips from patients with resistance to or failure to cross through the side of a stent there was tip damage. Even when the balloon crossed without perceptible resistance, tip damage occurred in over half of balloons 18/34 (53%). Some balloon designs were more resistant to damage than others. Tips from balloons of different nominal diameters from the same manufacturer had the same width. Stent distortion caused by damaged balloon tips is improved by kissing balloon post-dilatation. Balloon tip damage is common with crossing between stent struts. This is one cause of failure of a balloon to access a side-branch and a new balloon should be used. If stent distortion is suspected, it should be corrected with kissing balloon post-dilatation.Keywords
Funding Information
- Auckland Heart Group Charitable Trust
This publication has 8 references indexed in Scilit:
- Lessons from the real bench: non-BRSEuroIntervention, 2015
- Stent Longitudinal Strength Assessed Using Point CompressionCirculation: Cardiovascular Interventions, 2014
- Stent Longitudinal IntegrityJACC: Cardiovascular Interventions, 2011
- Randomized Comparison of Final Kissing Balloon Dilatation Versus No Final Kissing Balloon Dilatation in Patients With Coronary Bifurcation Lesions Treated With Main Vessel StentingCirculation, 2011
- Consensus from the 5th European Bifurcation Club meetingEuroIntervention, 2010
- Classification of coronary artery bifurcation lesions and treatments: Time for a consensus!Catheterization and Cardiovascular Interventions, 2007
- Optimal final lumen area and predictors of target lesion revascularization after stent implantation in small coronary arteries.The American Journal of Cardiology, 2003
- Immediate and one-year outcome in patients with coronary bifurcation lesions in the modern era (NHLBI dynamic registry)The American Journal of Cardiology, 2001