Experimental and numerical investigations of fracture and fatigue behaviour of implant-supported bars with distal extension made of three different materials
Authors:
Open Access
- 15 December 2020
- journal article
- research article
- Published by Walter de Gruyter GmbH in Biomedical Engineering / Biomedizinische Technik
- Vol. 66 (3) , 305-316
Abstract
The aim of this study was to investigate experimentally the fatigue and stability of three bar materials with distal extension at the molar region and to numerically analyse the biomechanical properties of the bar materials connected to overdentures in a patient individual model. A milled bar was designed for the mandible on four implants in the canine and second premolar region. Three bar materials were investigated: titanium (Ti), cobalt chromium (CoCr), and polyetherketonketon (PEKK). Firstly, static and fatigue tests were performed based on EN ISO 14801 in a commercial permanent loading set-up. Unilateral axial force was applied on the distal extension of the bars. Secondly, numerical models were created. Different bar materials and loading scenarios were analysed. The static fracture limit of the three materials was 1,750 N, 780 N, 310 N for Ti, CoCr, and PEKK, respectively. The Wöhler curves showed comparable fatigue limits of 200 N, 160 N, and 150 N for titanium, CoCr, and PEKK, respectively. The stress at the distal extension was 2,600 MPa (Ti), 1,000 MPa (CoCr), and 270 MPa (PEKK). All loading simulations with the PEKK bar showed higher stresses in the implants and in the bone bed as well as higher displacements of the over denture in comparison to metal bars. PEKK showed different mechanical behaviour compared to Ti and CoCr. The distribution of stresses within the PEKK bar was wider than the area of loading which probably leads to fatigue of the whole bar and not only the part under load.Keywords
This publication has 25 references indexed in Scilit:
- The Impact of Edentulism on Oral and General HealthInternational Journal of Dentistry, 2013
- Retention force of plastic clips on implant bars: a randomized controlled trialClinical Oral Implants Research, 2011
- Effects of Horizontal Misfit and Bar Framework Material on the Stress Distribution of an Overdenture-Retaining Bar System: A 3D Finite Element AnalysisJournal of Prosthodontics, 2011
- Influence of Material of Overdenture‐Retaining Bar with Vertical Misfit on Three‐Dimensional Stress DistributionJournal of Prosthodontics, 2010
- Attachment Systems for Implant Retained Overdentures: A Literature ReviewImplant Dentistry, 2006
- The impact of conventional and implant supported prostheses on social and sexual activities in edentulous adultsJournal of Dentistry, 2005
- A clinical trial of patient satisfaction and prosthodontic needs with ball and bar attachments for implant-retained complete overdentures: Three-year resultsThe Journal of Prosthetic Dentistry, 2005
- Finite element analysis of non‐axial versus axial loading of oral implants in the mandible of the dogJournal of Oral Rehabilitation, 1998
- In vitro changes in clips and bars used to retain implant overdenturesThe Journal of Prosthetic Dentistry, 1995
- Biomechanical considerations in osseointegrated prosthesesThe Journal of Prosthetic Dentistry, 1983