Abstract
A novel inverter-based digital floating-gate MOSFET sensor design for commercial X-ray dosimetry is presented. The biomedical healthcare industry sterilizes blood products for storage purposes using Gamma and X-ray radiations. This requires an ultra-low-power dosimeter that ensures irradiation does not exceed the maximum allowable 50 Gy while providing the required minimum levels of 25 Gy. In this work, minimum-sized MOS transistor devices are employed in an inverter configuration, eliminating the continuous flow of current and reducing power consumption significantly. Maximum measured currents, which flow only during the transition period, are in the nA range, compared to continuous currents of conventional sensor designs in the μA range. Final measured results show the viability of the proposed design for radiation dosimetry applications.Keywords
Funding Information
- Natural Sciences and Engineering Research Council of Canada
This publication has 13 references indexed in Scilit:
- Development of a 2.45 GHz Antenna for Flexible Compact Radiation Dosimeter TagsIEEE Transactions on Antennas and Propagation, 2019
- Floating Gate sensor for in-vivo dosimetry in radiation therapies. Design and first characterization.Journal of Physics: Conference Series, 2017
- Re-usable 180nm CMOS dosimeter based on a floating gate device2016
- Characterization and Modeling of a Floating Gate Dosimeter with Gamma and Protons at Various EnergiesIEEE Transactions on Nuclear Science, 2014
- Radiation sensor based on a floating gate device2012
- Ultra Low Power CMOS-Based Sensor for On-Body Radiation Dose MeasurementsIEEE Journal on Emerging and Selected Topics in Circuits and Systems, 2012
- Floating Gate CMOS Dosimeter With Frequency OutputIEEE Transactions on Nuclear Science, 2012
- Guidelines on the use of irradiated blood components prepared by the British Committee for Standards in Haematology blood transfusion task forceBritish Journal of Haematology, 2010
- Wireless Dosimeter: System-on-Chip Versus System-in-Package for Biomedical and Space ApplicationsIEEE Transactions on Circuits and Systems II: Express Briefs, 2008
- A sensitive, temperature-compensated, zero-bias floating gate MOSFET dosimeterIEEE Transactions on Nuclear Science, 2004