High resolution on a quadrupole ion trap mass spectrometer
Authors:
- 1 May 1991
- journal article
- Published by American Chemical Society (ACS) in Journal of the American Society for Mass Spectrometry
- Vol. 2 (3) , 198-204
Abstract
By using a modified ion trap mass spectrometer, resolution in excess of 30,000 (FWHM) at m I z 502 is demonstrated. The method of increasing resolution in the ion trap mass spectrometer operated in the mass-selective instability mode depends on decreasing the rate of scanning the primary radio frequency amplitude as well as using resonance ejection at the appropriate frequency and amplitude. A theoretical basis for the method is introduced.Keywords
This publication has 11 references indexed in Scilit:
- Electrospray ionization combined with ion trap mass spectrometryAnalytical Chemistry, 1990
- Ion isolation and sequential stages of mass spectrometry in a quadrupole ion trap mass spectrometerInternational Journal of Mass Spectrometry and Ion Processes, 1990
- Extending the mass range of the quadrupole ion trap using axial modulationRapid Communications in Mass Spectrometry, 1989
- Injection of ions into a quadrupole ion trap mass spectrometerInternational Journal of Mass Spectrometry and Ion Processes, 1989
- Mass range extension in a quadrupole ion‐trap mass spectrometerRapid Communications in Mass Spectrometry, 1989
- Theoretical foundation of molecular structures useful in structure/property relationshipsAnalytica Chimica Acta, 1988
- Ion/molecule reactions investigated by using tandem mass spectrometry in an ion trap: halomethylationAnalytica Chimica Acta, 1988
- Electrospray interface for liquid chromatographs and mass spectrometersAnalytical Chemistry, 1985
- Recent improvements in and analytical applications of advanced ion trap technologyInternational Journal of Mass Spectrometry and Ion Processes, 1984
- Die dreidimensionale Stabilisierung von Ladungstr gern in einem VierpolfeldThe European Physical Journal A, 1959