Breast tomography with synchrotron radiation: preliminary results
Authors:
- Silvia PaniSPSilvia Pani
- Renata LongoRLRenata Longo
- Diego DreossiDDDiego Dreossi
- Francesco MontanariFMFrancesco Montanari
- Alessandro OlivoAOAlessandro Olivo
- Fulvia ArfelliFAFulvia Arfelli
- Anna BergamaschiABAnna Bergamaschi
- Paolo PoropatPPPaolo Poropat
- Luigi RigonLRLuigi Rigon
- Fabrizio ZanconatiFZFabrizio Zanconati
- Ludovico Dalla PalmaLPLudovico Dalla Palma
- Edoardo CastelliECEdoardo Castelli
- 1 April 2004
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 49 (9) , 1739-1754
Abstract
A system for in vivo breast imaging with monochromatic x-rays has been designed and built at the synchrotron radiation facility Elettra in Trieste (Italy) and will be operational in 2004. The system design involves the possibility of performing both planar mammography and breast tomography. In the present work, the first results obtained with a test set-up for breast tomography are shown and discussed. Tomographic images of in vitro breasts were acquired using monochromatic x-ray beams in the energy range 20-28 keV and a linear array silicon pixel detector. Tomograms were reconstructed using standard filtered backprojection algorithms; the effect of different filters was evaluated. The attenuation coefficients of fibroglandular and adipose tissue were measured, and a quantitative comparison of images acquired at different energies was performed by calculating the differential signal-to-noise ratio of fibroglandular details in adipose tissue. All images required a dose comparable to the dose delivered in clinical, conventional mammography and showed a high resolution of the breast structures without the overlapping effects that limit the visibility of the structures in 2D mammography. A quantitative evaluation of the images proves that the image quality at a given dose increases in the considered energy range and for the considered breast sizes.Keywords
This publication has 28 references indexed in Scilit:
- High-speed single photon counting read out electronics for a digital detection system for clinical synchrotron radiation mammographyNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2004
- FROST: an ASIC for digital mammography with synchrotron radiationNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2003
- Dedicated Breast CT: Radiation Dose and Image Quality EvaluationRadiology, 2001
- Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocolPhysics in Medicine & Biology, 2000
- Mammography with Synchrotron Radiation: Phase-Detection TechniquesRadiology, 2000
- Low-dose phase contrast x-ray medical imagingPhysics in Medicine & Biology, 1998
- Mammography of a phantom and breast tissue with synchrotron radiation and a linear-array silicon detector.Radiology, 1998
- At the frontiers of digital mammography: SYRMEPNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1998
- Castor 1.0, a VLSI analog-digital circuit for pixel imaging applicationsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1997
- Digital mammography at the Trieste synchrotron light sourceIEEE Transactions on Nuclear Science, 1996