Enhancing the Specificity of the COBAS AMPLICOR CT/NG Test for Neisseria gonorrhoeae by Retesting Specimens with Equivocal Results
Authors:
Open Access
- 1 September 2001
- journal article
- clinical trial
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 39 (9) , 3092-3098
Abstract
The COBAS AMPLICOR CT/NG test for Neisseria gonorrhoeae cross-reacts with certain strains of nonpathogenic Neisseria species. In some strains, the target sequence is identical to that of N. gonorrhoeae , whereas other strains have a small number of mismatches within the regions recognized by the primers or probe used in the COBAS AMPLICOR NG test. These cross-reactive strains are occasionally present in urogenital specimens, causing false-positive results in the COBAS AMPLICOR NG test. Analysis of the data generated in a large multicenter clinical trial showed that 2.9% of the specimens gave signals between A 660 s of 0.2 and 3.5 but that one-half of these equivocal specimens did not contain N. gonorrhoeae . Most of these equivocal specimens were correctly classified as true positive or true negative by retesting in duplicate and defining a PCR-positive result as two of three results with an A 660 of ≥2.0. If specimens had been classified as positive or negative based on a single test result using a cutoff of an A 660 of 0.2, specificity would have ranged from 96.2 to 98.9% depending on specimen type, sex, and presence of symptoms. By employing the equivocal zone-retesting algorithm, specificity increased to 98.6 to 99.9% with little effect (0.1 to 4.9% decrease) on sensitivity in most specimen types, enabling the test to achieve a positive predictive value of at least 90% in populations with a prevalence of 4% or higher. In lower-prevalence populations, the test could be used to screen for presumptive infections that would have to be confirmed by an independent test.Keywords
This publication has 12 references indexed in Scilit:
- Evaluation of AMPLICOR Neisseria gonorrhoeae PCR Using cppB Nested PCR and 16S rRNA PCRJournal of Clinical Microbiology, 1999
- Evaluation of the Abbott LCx Assay for Detection of Neisseria gonorrhoeae in Endocervical Swab Specimens from FemalesJournal of Clinical Microbiology, 1998
- Detection of Neisseria gonorrhoeae Infection by Ligase Chain Reaction Testing of Urine Among Adolescent Women With and Without Chlamydia trachomatis InfectionSexually Transmitted Diseases, 1998
- Evaluation of the Abbott LCx Ligase Chain Reaction Assay for Detection of Chlamydia trachomatis and Neisseria gonorrhoeae in Urine and Genital Swab Specimens from a Sexually Transmitted Disease Clinic PopulationJournal of Clinical Microbiology, 1998
- Comparison of ligase chain reaction and culture for detection of Neisseria gonorrhoeae in genital and extragenital specimensJournal of Clinical Microbiology, 1997
- Detection of Chlamydia trachomatis and Neisseria gonorrhoeae by ligase chain reaction-based assays with clinical specimens from various sites: implications for diagnostic testing and screeningJournal of Clinical Microbiology, 1996
- MicroCorrespondenceMolecular Microbiology, 1996
- Ligase chain reaction for detection of Neisseria gonorrhoeae in urogenital swabsJournal of Clinical Microbiology, 1995
- Evaluation of ligase chain reaction for use with urine for identification of Neisseria gonorrhoeae in females attending a sexually transmitted disease clinicJournal of Clinical Microbiology, 1995
- A DNA sequence for the discrimination of Neisseria gonorrhoeae from other Neisseria speciesMolecular and Cellular Probes, 1991