Accuracy and Reproducibility of the Wound Shape Measuring and Monitoring System
Authors:
- Diane LangemoDLDiane Langemo
- James SpahnJSJames Spahn
- Linda SnodgrassLSLinda Snodgrass
- 1 July 2015
- journal article
- research article
- Published by Wolters Kluwer Health in Advances in Skin & Wound Care
- Vol. 28 (7) , 317-323
Abstract
Nineteen known area shapes simulated wounds and were measured using 3 methods and orientations (n = 57). Equivalence and accuracy were tested for ruler length × width and the WoundVision length × width and tracing functions. The WoundVision Coefficient of Individual Agreement value and accuracy were closest to the actual area (Psi_R = 0.77; 95% confidence interval, 0.528-1.016).Keywords
This publication has 3 references indexed in Scilit:
- Improving the Detection of Pressure Ulcers Using the TMI ImageMed SystemAdvances in Skin & Wound Care, 2011
- A Clinical Investigation into the Relationship between Increased Periwound Skin Temperature and Local Wound Infection in Patients with Chronic Leg UlcersAdvances in Skin & Wound Care, 2010
- Sizing up wounds accuratelyNursing, 2005