Accuracy of quantitative reconstructions in SPECT/CT imaging
Authors:
- S ShcherbininSSS Shcherbinin
- A CellerACA Celler
- T BelhocineTBT Belhocine
- R VanderwerfRVR Vanderwerf
- A DriedgerADA Driedger
- 4 August 2008
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 53 (17) , 4595-4604
Abstract
The goal of this study was to determine the quantitative accuracy of our OSEM-APDI reconstruction method based on SPECT/CT imaging for Tc-99m, In-111, I-123, and I-131 isotopes. Phantom studies were performed on a SPECT/low-dose multislice CT system (Infinia-Hawkeye-4 slice, GE Healthcare) using clinical acquisition protocols. Two radioactive sources were centrally and peripherally placed inside an anthropometric Thorax phantom filled with non-radioactive water. Corrections for attenuation, scatter, collimator blurring and collimator septal penetration were applied and their contribution to the overall accuracy of the reconstruction was evaluated. Reconstruction with the most comprehensive set of corrections resulted in activity estimation with error levels of 3-5% for all the isotopes.Keywords
This publication has 27 references indexed in Scilit:
- Quantification in emission tomography: Challenges, solutions, and performanceNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2007
- Hybrid scatter correction applied to quantitative holmium-166 SPECTPhysics in Medicine & Biology, 2006
- Model-based compensation for quantitative123I brain SPECT imagingPhysics in Medicine & Biology, 2006
- The Impact of PET and SPECT on Dosimetry for Targeted Radionuclide TherapyZeitschrift für Medizinische Physik, 2006
- Automated quality control of emission-transmission misalignment for attenuation correction in myocardial perfusion imaging with SPECT-CT systemsJournal of Nuclear Cardiology, 2006
- Validation of the Monte Carlo simulator GATE for indium-111 imagingPhysics in Medicine & Biology, 2005
- Dual-Modality Imaging of Cancer with SPECT/CTTechnology in Cancer Research & Treatment, 2002
- A slice-by-slice blurring model and kernel evaluation using the Klein-Nishina formula for 3D scatter compensation in parallel and converging beam SPECTPhysics in Medicine & Biology, 2000
- A scatter model for parallel and converging beam SPECT based on the Klein-Nishina formulaIEEE Transactions on Nuclear Science, 1994
- Inverse Monte Carlo: A Unified Reconstruction Algorithm for SPECTIEEE Transactions on Nuclear Science, 1985