Experimental validation of a simple model capable of predicting the phase contrast imaging capabilities of any x-ray imaging system
Authors:
- A OlivoAOA Olivo
- R SpellerRSR Speller
- 31 May 2006
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 51 (12) , 3015-3030
Abstract
Phase contrast (PC) imaging is one of the most exciting emerging x-ray imaging techniques, with the potential of removing some of the main limitations of conventional radiology. After extensive experimentation carried out particularly at synchrotron radiation (SR) facilities, the scientific community agrees that it is now time to translate these ideas towards the first clinical implementations. In this framework, a complete model, based on Fresnel/Kirchoff diffraction integrals, was devised. This model accounts for source dimensions, beam spectrum and divergence and detector point spread function (PSF), and can thus be applied to any x-ray imaging system. In particular, by accepting in input the above parameters along with the ones describing the sample, the model can be used to optimize the geometry of the set-up, i.e. to assess the source-to-sample and sample-to-detector distances which maximize feature detection. The model was evaluated by acquiring a range of images of different samples with a laboratory source, and a good agreement was found between simulated and experimental data in all cases. In order to maximize the generality of the results, all acquisitions were carried out using a polychromatic source and an energy-resolving detector; in this way, a range of monochromatic images could be obtained as well as polychromatic images, which can be created by integrating different parts of the acquired spectra. One of the most notable results obtained is that in many practical cases polychromatic PC imaging can provide the same image quality as its monochromatic counterpart. This is an important step in the wider application of PC using conventional sources.Keywords
This publication has 24 references indexed in Scilit:
- Experimental validation of the Wigner distributions theory of phase‐contrast imagingMedical Physics, 2005
- Quantification of the effect of kVp on edge‐enhancement index in phase‐contrast radiographyMedical Physics, 2002
- Quantitative In-Line Phase-Contrast Imaging with Multienergy X RaysPhysical Review Letters, 2001
- X-ray emission from a compact hot plasma: applications to radiology and mammographyPhysics in Medicine & Biology, 2001
- Mammography with Synchrotron Radiation: Phase-Detection TechniquesRadiology, 2000
- Holotomography: Quantitative phase tomography with micrometer resolution using hard synchrotron radiation x raysApplied Physics Letters, 1999
- Low-dose phase contrast x-ray medical imagingPhysics in Medicine & Biology, 1998
- Diffraction enhanced x-ray imagingPhysics in Medicine & Biology, 1997
- X-ray plane-wave topography observation of the phase contrast from a non-crystalline objectJournal of Physics D: Applied Physics, 1995
- Phase-contrast imaging of weakly absorbing materials using hard X-raysNature, 1995