Background/purpose: This study has evaluated a multi‐parametric system combining laser Doppler flowmetry and photoplethysmography in a single probe for the simultaneous measurement of blood flow at different depths in the tissue. This system will be used to facilitate the understanding of pressure ulcer formation and in the evaluation of pressure ulcer mattresses.Methods: The blood flow in the tissue over the sacrum was measured before, during and after loading with 37.5 mmHg, respectively, 50.0 mmHg. The evaluation of the system consisted of one clinical part, and the other part focusing on the technicalities of the probe prototype.Results: An increase in blood flow while loading was the most common response, but when the blood flow decreased during loading it was most affected at the skin surface and the blood flow responses may be different due to depths of measurement. Reactive hyperaemia may occur more frequently in the superficial layers of the tissue.Conclusion: The study showed that the new system is satisfactory for measuring tissue blood flow at different depths. The laser Doppler complements the photoplethysmography, and further development of the system into a thin flexible probe with the ability to measure a larger area is required.Keywords
This publication has 26 references indexed in Scilit:
- Photoplethysmography and its application in clinical physiological measurementPhysiological Measurement, 2007
- Pressure ulcers and their treatment and effects on quality of life: hospital inpatient perspectivesJournal of Advanced Nursing, 2007
- Patient stories of living with a pressure ulcerJournal of Advanced Nursing, 2006
- Non-invasive methods and stimuli for evaluating the skin's microcirculationJournal of Pharmacological and Toxicological Methods, 2006
- A Comparison of Changes in Rhythms of Sacral Skin Blood Flow in Response to Heating and IndentationArchives of Physical Medicine and Rehabilitation, 2005
- Non‐invasive monitoring of muscle blood perfusion by photoplethysmography: evaluation of a new applicationActa Physiologica Scandinavica, 2005
- Dynamics of Local Pressure-Induced Cutaneous Vasodilation in the Human HandMicrovascular Research, 2001
- Dynamic Coherence Analysis of Vasomotion and Flow Motion in Skeletal Muscle MicrocirculationMicrovascular Research, 1996
- Optical properties of blood in motionOptical Engineering, 1993
- Matching patient need for pressure sore prevention with the supply of pressure redistributing mattressesJournal of Advanced Nursing, 1992