An improved method for flat‐field correction of flat panel x‐ray detector
Authors:
- J. Anthony SeibertJSJ. Anthony Seibert
- John M. BooneJBJohn M. Boone
- 24 January 2006
- journal article
- Published by Wiley in Medical Physics
- Vol. 33 (2) , 391-393
Abstract
In this Technical Note, the effects of different flat‐field techniques are examined for a cesium iodide flat panel detector, which exhibited a slightly nonlinear exposure response. The results indicate that the variable flat‐field correction method with the appropriate polynomial fit provides excellent correction throughout the entire exposure range. The averaged normalized variation factor, used to assess the nonuniformity of the flat‐field correction, decreased from 30.76 for the fixed correction method to 4.13 for the variable flat‐field correction method with a fourth‐order polynomial fit for the spectrum, and from 16.42 to 3.97 for the spectrum.Keywords
Funding Information
- California Breast Cancer Research Program (7EB‐0075)
- National Cancer Institute (CA‐89260)
This publication has 4 references indexed in Scilit:
- Progress in the development of a dedicated breast CT scanner2004
- A ghost story: Spatio-temporal response characteristics of an indirect-detection flat-panel imagerMedical Physics, 1999
- How does real offset and gain correction affect the DQE in images from x-ray flat detectors?1999
- Flat-field correction technique for digital detectors1998