Utility of Intraoperative Radiographs in Pediatric Cochlear Implant Surgery
- 13 April 2023
- journal article
- research article
- Published by Wiley in Otolaryngology–Head and Neck Surgery
- Vol. 169 (5) , 1282-1289
Abstract
Objective: To evaluate the role of intraoperative radiographs to confirm electrode position following pediatric cochlear implantation (CI).Study Design: Retrospective chart review.Setting: Single tertiary care pediatric center.Methods: A retrospective chart review was conducted, including all pediatric patients undergoing CI at UPMC Children's Hospital of Pittsburgh over a 13‐year period.Results: We identified 326 patients undergoing 492 procedures. Across the cohort, there were 7 cases that required intraoperative electrode reinsertion due to malposition or presumed malposition. For 6 of the 7 cases, intraoperative X‐ray identified electrode malposition. Neural response telemetry (NRT) testing was also abnormal for 4 of these cases prior to reinsertion. Implantation of Cochlear's Slim Modiolar electrode was associated with an abnormal perioperative X‐ray (odds ratio [OR]: 9.2, p = 0.03) and increased change in management (OR: 9.2, p = 0.03) compared to Cochlear's Contour Advance (CA). Incidence of abnormal X‐rays was 1.24% overall, 4% in the Slim Modiolar group, and 0.3% in the CA group. The Slim Modiolar electrode accounted for 4 of 7 cases requiring reinsertion, and in all 4 of these cases, electrode fold‐over was identified on the X‐ray. NRT was normal in 1 of these 4 cases.Conclusion: The use of Cochlear's Slim Modiolar electrode was associated with a significantly increased risk of abnormal intraoperative X‐ray compared to the CA electrode. Given the risk of fold‐over with routine insertion and normal electrical testing using the Slim Modiolar electrode, we recommend routine use of intraoperative skull X‐ray to confirm electrode position.Keywords
This publication has 18 references indexed in Scilit:
- Scalar localisation of peri-modiolar electrodes and speech perception outcomesThe Journal of Laryngology & Otology, 2018
- Incidence for Tip Foldover During Cochlear ImplantationOtology & Neurotology, 2018
- Early Outcomes With a Slim, Modiolar Cochlear Implant Electrode ArrayOtology & Neurotology, 2018
- Utility of intraoperative computed tomography for cochlear implantation in patients with difficult anatomyCochlear Implants International, 2017
- Utility of intraoperative and postoperative radiographs in pediatric cochlear implant surgeryInternational Journal of Pediatric Otorhinolaryngology, 2017
- Intra-operative radiological diagnosis of a tip roll-over electrode array displacement using fluoroscopy, when electrophysiological testing is normal: the importance of both techniques in cochlear implant surgeryBrazilian Journal of Otorhinolaryngology, 2017
- Intra-operative skull X-ray for misdirection of the cochlear implant array into the vestibular labyrinthThe Journal of Laryngology & Otology, 2015
- Cochlear implant electrode misplacement: Incidence, evaluation, and managementThe Laryngoscope, 2013
- An Evidence-Based Algorithm for Intraoperative Monitoring During Cochlear ImplantationOtology & Neurotology, 2012
- Real-Time Intraoperative Computed Tomography to Assist Cochlear Implant Placement in the Malformed Inner EarOtology & Neurotology, 2009