Validation of Continuous Noninvasive Blood Pressure Monitoring Using Error Grid Analysis

Authors:

Abstract
BACKGROUND: Error grid analysis was recently proposed to compare blood pressure obtained by 2 measurement methods. This study aimed to compare continuous noninvasive blood pressure (CNBP) with invasive blood pressure (IBP) using the error grid analysis and investigate the confounding risk factors attributable to the differences between CNBP and IBP. METHODS: Sixty adult patients undergoing general anesthesia were prospectively enrolled. Simultaneous comparative data regarding CNBP and IBP were collected. The Bland-Altman analysis was conducted to compare CNBP and IBP for systolic blood pressure (SBP) and mean blood pressure (MBP; acceptable accuracy: mean bias P = .004) and length of time from the initiation of finger cuff inflation (adjusted odds ratio for 60 minutes: 1.40, 95% CI, 1.13–1.73; P = .002) were significant factors of being in a more dangerous zone of the error grid. CONCLUSIONS: The error grid analysis revealed the larger clinical discrepancy between CNBP and IBP in MBP compared with that in SBP. Old age and longer finger cuff inflation time were significant factors of being in a more dangerous zone of the error grid, which could affect the hemodynamic management during surgery.

This publication has 26 references indexed in Scilit: