Abstract
This work presents the development of an intelligent microsurgical instrument to perform real-time tremor compensation within a handheld tool. The intelligent instrument senses its own motion, distinguishes between voluntary and erroneous motion, and manipulates its tip to cancel the undesired component in real-time. The on-board sensing unit is made up of a magnetometer-aided all-accelerometer inertial measurement unit and sensor fusion is performed via a quaternion-based Kalman filtering. Tremor is modeled and filtered by an adaptive zero-phase notch filter. The intraocular shaft manipulator is a three DOF piezoelectric actuated mechanism driven by a feedforward controller with inverse rate-dependent hysteresis model. Laboratory experimental results of the system are presented.Keywords
This publication has 12 references indexed in Scilit:
- Precision measurement for microsurgical instrument evaluation2005
- Physiological tremor amplitude during retinal microsurgery2002
- A Study of Surgical Approaches to Retinal Vascular OcclusionsArchives of Ophthalmology (1950), 2000
- Evaluation of a Telerobotic System to Assist Surgeons in MicrosurgeryComputer Aided Surgery, 1999
- Adaptive cancelling of physiological tremor for improved precision in microsurgeryIEEE Transactions on Biomedical Engineering, 1998
- Quantitative Measurement of the Effects of Caffeine and Propranolol on Surgeon Hand TremorArchives of Ophthalmology (1950), 1997
- Prevalence and Associations of Retinal Vein Occlusion in AustraliaArchives of Ophthalmology (1950), 1996
- A Teleoperated Microsurgical Robot and Associated Virtual Environment for Eye SurgeryPRESENCE: Virtual and Augmented Reality, 1993
- Physiologic tremor and microsurgeryMicrosurgery, 1983
- ERGONOMICS APPLIED TO THE PRACTICE OF MICROSURGERY1Anz Journal of Surgery, 1977