Impact of a self-developed planning and self-constructed navigation system on skull base surgery: 10 years experience
- 1 January 2007
- journal article
- research article
- Published by Taylor & Francis in Acta Oto-Laryngologica
- Vol. 127 (4) , 403-407
Abstract
Conclusion. Our self-developed planning and navigation system has proven its capacity for accurate surgery on the anterior and lateral skull base. With the incorporation of augmented reality, image-guided surgery will evolve into 'information-guided surgery’. Objective. Microscopic or endoscopic skull base surgery is technically demanding and its outcome has a great impact on a patient's quality of life. The goal of the project was aimed at developing and evaluating enabling navigation surgery tools for simulation, planning, training, education, and performance. This clinically applied technological research was complemented by a series of patients (n=406) who were treated by anterior and lateral skull base procedures between 1997 and 2006. Materials and methods. Optical tracking technology was used for positional sensing of instruments. A newly designed dynamic reference base with specific registration techniques using fine needle pointer or ultrasound enables the surgeon to work with a target error of < 1 mm. An automatic registration assessment method, which provides the user with a color-coded fused representation of CT and MR images, indicates to the surgeon the location and extent of registration (in)accuracy. Integration of a small tracker camera mounted directly on the microscope permits an advantageous ergonomic way of working in the operating room. Additionally, guidance information (augmented reality) from multimodal datasets (CT, MRI, angiography) can be overlaid directly onto the surgical microscope view. The virtual simulator as a training tool in endonasal and otological skull base surgery provides an understanding of the anatomy as well as preoperative practice using real patient data. Results. Using our navigation system, no major complications occurred in spite of the fact that the series included difficult skull base procedures. An improved quality in the surgical outcome was identified compared with our control group without navigation and compared with the literature. The surgical time consumption was reduced and more minimally invasive approaches were possible. According to the participants’ questionnaires, the educational effect of the virtual simulator in our residency program received a high ranking.Keywords
This publication has 9 references indexed in Scilit:
- Navigational Risks Associated With Sinus Surgery and the Clinical Effects of Implementing a Navigational System for Sinus SurgeryThe Laryngoscope, 2009
- Nationwide Survey on the use of Image‐Guided Functional Endoscopic Sinus SurgeryOtolaryngology–Head and Neck Surgery, 2006
- Improved targeting device and computer navigation for accurate placement of brachytherapy needlesMedical Physics, 2005
- A-Mode Ultrasound–Based Registration in Computer-Aided Surgery of the SkullJAMA Otolaryngology–Head & Neck Surgery, 2003
- Valuable use of computer-aided surgery in congenital bony aural atresiaThe Journal of Laryngology & Otology, 2003
- Image‐Guided Sinus Surgery: Lessons Learned from the First 1000 CasesOtolaryngology–Head and Neck Surgery, 2003
- Frameless Optical Computer-Aided Tracking of a Microscope for Otorhinology and Skull Base SurgeryJAMA Otolaryngology–Head & Neck Surgery, 2001
- Management of patients in fast track surgeryBMJ, 2001
- Konzept eines rahmenlosen bildinteraktiven Navigations-systems für die Schädelbasis-, Nasen- und Nasennebenhöhlen-chirurgieOto-Rhino-Laryngologia Nova, 1997