Evaluation of the structural behavior of three and four implant-supported fixed prosthetic restorations by finite element analysis
Authors:
- Santiago CorreaSCSantiago Correa
- Juliana IvancikJIJuliana Ivancik
- Juan Felipe IsazaJIJuan Felipe Isaza
- Mauricio NaranjoMNMauricio Naranjo
- 1 April 2012
- journal article
- Published by Japan Prosthodontic Society in Journal of Prosthodontic Research
- Vol. 56 (2) , 110-119
Abstract
There is much controversy about the minimum number of implants and maximum cantilever length in mandible prosthetic restoration. Finite elements analysis of three and four implant-supported prostheses was performed to determine the stresses in the superstructure, implants and cortical bone and, therefore, the failure prediction for each restoration. An edentulous mandible was modeled from CT scan images. Two finite element models of three and four implant-supported prostheses with cantilever lengths of 10 and 15 mm were created. Occlusal loads in different parts of the superstructure were applied and shear and normal stresses were calculated. Two failure criteria were analyzed: the von Mises criterion for isotropic materials (superstructure and implants) and the Tsai-Wu criterion for transversely isotropic material (cortical bone). Both criteria predict failure in the three implant-supported prosthesis for all cases analyzed. The same applies for the four-implant prosthesis of 15 mm cantilever length. However, four implants and a cantilever length of 10mm passed the failure criteria and were considered safe. The results from the patient analyzed showed that fixed support prostheses on three implants are not recommended from a structural point of view because they do not adequately support occlusal loads. Excessive stress in the superstructure and the cortical bone can be expected, which would anticipate the failure of the restoration. Fixed support prostheses on four implants with a cantilever length of 10mm properly resist occlusal loading.Keywords
This publication has 32 references indexed in Scilit:
- Influence of maxillary cortical bone thickness, implant design and implant diameter on stress around implants: A three-dimensional finite element analysisJournal of Prosthodontic Research, 2010
- Selections of the cylinder implant neck taper and implant end fillet for optimal biomechanical properties: A three-dimensional finite element analysisJournal of Biomechanics, 2008
- “All‐on‐Four” Immediate‐Function Concept with Brånemark System® Implants for Completely Edentulous Mandibles: A Retrospective Clinical StudyClinical Implant Dentistry and Related Research, 2003
- Comparative evaluation of the effect of diameter, length and number of implants supporting three-unit fixed partial prostheses on stress distribution in the boneJournal of Dentistry, 2002
- Mechanical properties of cancellous bone in the human mandibular condyle are anisotropicJournal of Biomechanics, 2001
- Cantilever and Implant Biomechanics: A Review of the Literature, Part 2Journal of Prosthodontics, 1994
- Elastic properties of human supraorbital and mandibular boneAmerican Journal of Physical Anthropology, 1993
- Three‐dimensional finite element stress analysis of the dentate human mandibleAmerican Journal of Physical Anthropology, 1992
- Edentulousness—an oral handicap. Patient reactions to treatment with iawbone‐anchored prosthesesJournal of Oral Rehabilitation, 1987
- A General Theory of Strength for Anisotropic MaterialsJournal of Composite Materials, 1971