Effects of Internal Filtration on the Solute Removal Efficiency of a Dialyzer
Authors:
- Isamu IshimoriIIIsamu Ishimori
- Kazuhiro IshidaKIKazuhiro Ishida
- Toshihisa HoshinoTHToshihisa Hoshino
- Iwakazu KanekoIKIwakazu Kaneko
- Yuichi SatoYSYuichi Sato
- Tetsuzo AgishiTATetsuzo Agishi
- Noriyuki TamamuraNTNoriyuki Tamamura
- Hidehiko SakuraiHSHidehiko Sakurai
- Toshiaki MasudaTMToshiaki Masuda
- Hiroyuki HattoriHHHiroyuki Hattori
- 1 July 2000
- journal article
- research article
- Published by Wolters Kluwer Health in ASAIO Journal
- Vol. 46 (4) , 456-460
Abstract
To improve solute removal efficiency, several types of dialyzers with enhanced internal filtration were introduced for clinical application. In these dialyzers, enhanced internal filtration increased convective transport of the solute, in addition to diffusive transport. In this study, the effects of internal filtration on solute removal efficiency were examined by both analytic and experimental studies. Internal filtration is affected by blood (Q B) and dialysate (Q D) flow rates; the patient’s hematocrit and plasma level of total protein; and the effective length (L eff), inner diameter (D), and density ratio (DR) of the hollow fibers. An analytic model was introduced for the estimation of the changes in mass and momentum along the dialyzer. It clarified the effects of these parameters on maximum internal filtration flow rate (Q IF) and clearance (K) of urea (60 daltons), vitamin B12 (1,355), and myoglobin (17,000). As a result of the analytic study, Q IF was increased, resulting in a smaller D, a longer L eff, and a larger DR value. Several types of dialyzers with the same cellulose triacetate membrane, produced by Toyobo Co, Ltd., Ohtsu, Japan, and Nissho Corporation, Kusatsu, Japan, were used for the experimental study. An in vitro evaluation using myoglobin solution showed the same trends as found in the analytic study. For example, a dialyzer with 150 μm of D has a 72.0 ml/min myoglobin K value, much higher than that of 53.7 ml/min for a dialyzer with 200 μm of D under constant Q B (300 ml/min) and DR (50%) values. Development of a dialyzer with enhanced internal filtration, however, should take the patient’s safety into account, and hemolysis and endotoxin invasion from the dialysate to the patient should be avoided.Keywords
This publication has 2 references indexed in Scilit:
- Internal filtration--advantage in haemodialysis?Nephrology Dialysis Transplantation, 1996
- Bulk Mass Transport Limitations during High‐Flux HemodialysisArtificial Organs, 1993