Improved crossability with novel cutting balloon versus scoring balloon in the treatment of calcified lesion
- 28 March 2020
- journal article
- research article
- Published by Springer Nature in Cardiovascular Intervention and Therapeutics
- Vol. 36 (2) , 198-207
Abstract
Cutting balloons and scoring balloons are commonly used for the preparation of calcified lesion. However, problems with crossability occasionally limit the use of cutting balloons. We prospectively selected 173 calcified lesions treated using a novel cutting balloon (Wolverine™, C group). As control, we retrospectively analyzed 146 calcified lesions treated using a scoring balloon (Lacrosse NSE ALPHA™, S group). Either intravascular ultrasound (IVUS) or optical coherence tomography (OCT) was used by the operator’s discretion. The primary outcome was delivery success, which was defined as successful passage to the target lesion immediately after IVUS or OCT evaluation. The secondary outcome was acute cross-sectional area (CSA) gain, which was defined as post-interventional minimum stent area minus pre-procedural minimum lumen area. A multivariate analysis evaluated the independent predictors for delivery success. The delivery success rate was significantly higher in the C group versus the S group (90.8% versus 79.5%, P = 0.006). However, the acute CSA gain was similar between the two groups (IVUS: 3.2 ± 1.8 mm2 versus 3.4 ± 1.9 mm2, P = 0.53; OCT: 3.6 ± 1.4 mm2 versus 4.1 ± 1.9 mm2, P = 0.11). Usage of cutting balloon was an independent predictor for delivery success even after the adjustment for the patient and lesion characteristics [odds ratio (OR): 2.72 (95% confidence interval 1.38–5.33), P = 0.004] as well as the procedural characteristics [OR: 2.34 (1.15–4.86), P = 0.018]. Novel cutting balloons demonstrated better crossability and similar acute CSA gain compared with scoring balloons in calcified lesion.Keywords
This publication has 11 references indexed in Scilit:
- Effect of cutting balloon after rotational atherectomy in severely calcified coronary artery lesions as assessed by optical coherence tomographyCatheterization and Cardiovascular Interventions, 2019
- Intimal disruption affects drug-eluting cobalt-chromium stent expansion: A randomized trial comparing scoring and conventional balloon predilationInternational Journal of Cardiology, 2016
- The use of a scoring balloon for optimal lesion preparation prior to bioresorbable scaffold implantation: a comparison with conventional balloon predilatationEuroIntervention, 2016
- Consensus Standards for Acquisition, Measurement, and Reporting of Intravascular Optical Coherence Tomography StudiesJACC, 2012
- Impact of Post-Intervention Minimal Stent Area on 9-Month Follow-Up Patency of Paclitaxel-Eluting StentsJACC: Cardiovascular Interventions, 2009
- Impact of final stent dimensions on long-term results following sirolimus-eluting stent implantationJACC, 2004
- Sirolimus-Eluting Stents versus Standard Stents in Patients with Stenosis in a Native Coronary ArteryNew England Journal of Medicine, 2003
- Intravascular Ultrasound Analysis of Infarct-Related and Non–Infarct-Related Arteries in Patients Who Presented With an Acute Myocardial InfarctionCirculation, 2003
- American College of Cardiology clinical expert consensus document on standards for acquisition, measurement and reporting of intravascular ultrasound studies (ivus)31When citing this document, the American College of Cardiology would appreciate the following citation format: Mintz GS, Nissen SE, Anderson WD, Bailey SR, Erbel R, Fitzgerald PJ, Pinto FJ, Rosenfield K, Siegel RJ, Tuzcu EM, Yock PG. ACC Clinical Expert Consensus Document on Standards for the acquisition, measurement and reporting of intravascular ultrasound studies: a report of the American College of Cardiology Task Force on Clinical Expert Consensus Documents (Committee to Develop a Clinical Expert Consensus Document on Standards for Acquisition, Measurement and Reporting of Intravascular Ultrasound Studies [IVUS]). J Am Coll Cardiol2001;37:1478–92.33Address for reprints: This document is available on the ACC Website at www.acc.org. Reprints of this document are available for $5.00 each by calling 800-253-4636 (U.S. only) or writing to the Resource Center, American College of Cardiology, Educational Services, 9111 Old Georgetown Road, Bethesda, MD 20814-1699.JACC, 2001
- Experimental validation of geometric and densitometric coronary measurements on the new generation cardiovascular angiography analysis system (caas ii)Catheterization and Cardiovascular Diagnosis, 1993