A new dual threshold time-over-threshold circuit for fast timing in PET
Authors:
- Craig S LevinCLCraig S Levin
- 3 June 2014
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 59 (13) , 3421-3430
Abstract
Time-over-threshold (ToT) is attractive as a method of combined timing and energy encoding in positron emission tomography (PET) due to its simplicity in implementation and readout. However, conventional single threshold ToT has a nonlinear response and generally suffers from a tradeoff between timing and energy resolution. The resulting poor performance is not fit for applications requiring fast timing, such as time-of-flight (ToF) PET. In this work it is shown experimentally that by replacing single threshold ToT with a dual threshold method in a new compact circuit, excellent time resolution can be achieved (154 ps FWHM for 3 × 3 × 5 mm(3) LYSO crystals), suitable for ToF. Dual threshold ToT timing results have been compared to a similar single threshold design, demonstrating that dual threshold ToT performance is far superior to that of single threshold ToT (154 ps versus 418 ps coincidence time resolution for the dual and single threshold cases, respectively). A method of correcting for nonlinearity in dual threshold ToT energy spectra is also demonstrated.Keywords
This publication has 14 references indexed in Scilit:
- Design of a Real-Time FPGA-Based Data Acquisition Architecture for the LabPET II: An APD-Based Scanner Dedicated to Small Animal PET ImagingIEEE Transactions on Nuclear Science, 2013
- TOFPET ASIC for PET applicationsJournal of Instrumentation, 2013
- A new pulse width signal processing with delay-line and non-linear circuit (for ToT)Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2011
- Time-Based Readout of a Silicon Photomultiplier (SiPM) for Time of Flight Positron Emission Tomography (TOF-PET)IEEE Transactions on Nuclear Science, 2011
- Improvement in Lesion Detection with Whole-Body Oncologic Time-of-Flight PETJournal of Nuclear Medicine, 2011
- Multi-Level Time-Over-Threshold Method for Energy Resolving Multi-Channel SystemsIEEE Transactions on Nuclear Science, 2010
- A multi-threshold sampling method for TOF-PET signal processingNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2009
- Pulse width modulation: A novel readout scheme for high energy photon detection2008
- Benefit of Time-of-Flight in PET: Experimental and Clinical ResultsJournal of Nuclear Medicine, 2008
- A time-over-threshold machine: the readout integrated circuit for the BABAR Silicon Vertex TrackerIEEE Transactions on Nuclear Science, 1997