A new method for x-ray scatter correction: first assessment on a cone-beam CT experimental setup
Authors:
- L GerfaultLGL Gerfault
- F EstèveFEF Estève
- J-M DintenJDJ-M Dinten
- 17 July 2007
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 52 (15) , 4633-4652
Abstract
Cone-beam computed tomography (CBCT) enables three-dimensional imaging with isotropic resolution and a shorter acquisition time compared to a helical CT scanner. Because a larger object volume is exposed for each projection, scatter levels are much higher than in collimated fan-beam systems, resulting in cupping artifacts, streaks and quantification inaccuracies. In this paper, a general method to correct for scatter in CBCT, without supplementary on-line acquisition, is presented. This method is based on scatter calibration through off-line acquisition combined with on-line analytical transformation based on physical equations, to adapt calibration to the object observed. The method was tested on a PMMA phantom and on an anthropomorphic thorax phantom. The results were validated by comparison to simulation for the PMMA phantom and by comparison to scans obtained on a commercial multi-slice CT scanner for the thorax phantom. Finally, the improvements achieved with the new method were compared to those obtained using a standard beam-stop method. The new method provided results that closely agreed with the simulation and with the conventional CT scanner, eliminating cupping artifacts and significantly improving quantification. Compared to the beam-stop method, lower x-ray doses and shorter acquisition times were needed, both divided by a factor of 9 for the same scatter estimation accuracy.Keywords
This publication has 23 references indexed in Scilit:
- An alternate line erasure and readout (ALER) method for implementing slot-scan imaging technique with a flat-panel detector-initial experiencesIEEE Transactions on Medical Imaging, 2006
- TU‐D‐I‐611‐05: Monte Carlo Investigation of Scatter Contribution to Kilovoltage Cone‐Beam Computed Tomography ImagesMedical Physics, 2005
- Effect of scattered radiation on image noise in cone beam CTMedical Physics, 2001
- A slice-by-slice blurring model and kernel evaluation using the Klein-Nishina formula for 3D scatter compensation in parallel and converging beam SPECTPhysics in Medicine & Biology, 2000
- Regularization method for scatter‐glare correction in fluoroscopic imagesMedical Physics, 1999
- Extraction of primary signal from EPIDs using only forward convolutionMedical Physics, 1997
- 3-D image reconstruction with veiling glare correction to improve the contrast of 3-D reconstructed vascular imagesIEEE Transactions on Nuclear Science, 1996
- X‐ray scatter in megavoltage transmission radiography: Physical characteristics and influence on image qualityMedical Physics, 1994
- Monte Carlo simulation of the scattered radiation distribution in diagnostic radiologyMedical Physics, 1988
- Scattered radiation in fan beam imaging systemsMedical Physics, 1982