Objective. To objectively compare and analyze the range of motion of three types of instruments for single-incision laparoscopic surgery. Material and Methods. Ten experienced participants were recruited. Straight instruments (Group A), straight/articulating instruments (Group B), and precurved instruments (Group C) were used to complete the transferring task through one site in a laparoscopic simulator. Straight instruments via two separate sites (Group D) served as control. The operation time of each group was recorded. Instrument positions were measured by an optical tracking system. The inserted length and pivoting angles were derived via MATLAB. Results. There was a significant difference in operation time between groups ( , ). The range of motion of instruments was different on instrument types and surgical approaches. A significant difference in the inserted length was found between groups. Instrument conflicts and inadequate triangulation were found in Group A; instrument conflicts were found in Group B; no obvious conflicts and triangulation problems were observed in Group C. The operation in Group C was similar to the operation in Group D but differed on the left/right pivoting angles. Conclusion. Different types of instruments have different ranges of motion in single-incision laparoscopic surgery. Working with precurved instruments seems like a compromise to traditional laparoscopic surgery if the transmission property, and shaft curvature of the instruments could be improved. An integrated mechanical platform or robotic system might be the ultimate solution for single-incision laparoscopic surgery to pursue even less trauma.Keywords
Funding Information
- Second Military Medical University (2016QN16, 2019YFC0120402, 51901137, 51735003)
- Ministry of Science and Technology (2016QN16, 2019YFC0120402, 51901137, 51735003)
- National Natural Science Foundation of China (2016QN16, 2019YFC0120402, 51901137, 51735003)
This publication has 22 references indexed in Scilit:
- Comparison of endoscopic task performance with crossed versus uncrossed straight and curved instruments through a single portSurgical Endoscopy, 2012
- Pre-bent instruments used in single-port laparoscopic surgery versus conventional laparoscopic surgery: comparative study of performance in a dry labSurgical Endoscopy, 2012
- Effect of training and instrument type on performance in single-incision laparoscopy: Results of a randomized comparison using a surgical simulatorSurgical Endoscopy, 2011
- A Triangulating Operating Platform Enhances Bimanual Performance and Reduces Surgical Workload in Single-Incision LaparoscopyJournal of the American College of Surgeons, 2011
- Minimally invasive single-site surgery for the digestive system: A technological reviewJournal of Minimal Access Surgery, 2011
- Comparative Assessment of Laparoscopic Single-Site Surgery Instruments to Conventional Laparoscopic in Laboratory SettingJournal of Endourology, 2010
- Single-port laparoscopic surgery: an overviewSurgical Endoscopy, 2009
- Real-time 3D motion tracking for small animal brain PETPhysics in Medicine & Biology, 2008
- Ergonomics with the use of curved versus straight laparoscopic graspers during rectosigmoid resection: results of a multiprofile comparative studySurgical Endoscopy, 2007
- Development of a Model for Training and Evaluation of Laparoscopic Skills 11This work was supported by an educational grant from United States Surgical Corporation (Auto Suture Canada).The American Journal of Surgery, 1998