Abstract
The ability of experienced anaesthetists to discern oxygen saturation by listening to the tones of a Datex AS3 pulse oximeter was examined. Five-second samples were recorded using a high fidelity patient simulator and replayed singly and in pairs. Whilst the lower saturations were generally recognized as lower, the perceived range was greatly compressed. Median perceived estimates for 70% saturation was 89%, for 80% was 93% and for 94% was 94%. When comparing pairs of samples, the direction of the difference was correctly discerned by 70% of anaesthetists for differences of 2%, rising to 95% for differences of greater than 8% oxygen saturation. The magnitude of the difference was consistently underestimated. With an actual difference of 20%, the median estimate was 5%. The results indicate that while qualitative estimate changes in oxygen saturation are moderately reliable, quantitative estimation is severely limited by a compromised perceived scale. This may lead to underestimation of the severity if the auditory signal is relied on in isolation. A non-linear (musical) scale may prove more appropriate and should be investigated. Testing experienced anaesthetists demonstrated that most could detect the direction, but not the magnitude of a change in saturation by listening to the change in pitch of a Datex AS3 pulse oximeter tone.Keywords
This publication has 7 references indexed in Scilit:
- Current equipment alarm sounds: friend or foe?Canadian Journal of Anesthesia/Journal canadien d'anesthésie, 2003
- Continuous auditory monitoring—how much information do we register?British Journal of Anaesthesia, 1999
- Response Times to Visual and Auditory Alarms during AnaesthesiaAnaesthesia and Intensive Care, 1996
- Hearing Acuity of Anesthesiologists and Alarm DetectionAnesthesiology, 1994
- Audibility and identification of auditory alarms in the operating room and intensive care unitErgonomics, 1993
- Can people hear the pitch change on a variable-pitch pulse oximeter?Journal of Clinical Monitoring and Computing, 1992
- Ergonomics: Anaesthetists’ use of auditory alarms in the operating roomJournal of Clinical Monitoring and Computing, 1985