Abstract
The long term behavior of total hip arthroplasty ultimately is determined by adaptive bone remodeling. Stress adaptation creating various degrees of metaphyseal bone atrophy has occurred with prostheses designed for tight distal fit and fill. This phenomenon has been explained as bone adaptation resolving the conflict between femoral component stiffness and bone flexibility. However, clinical observations indicate that a different premise may be operational when prostheses with tapered geometries are used. The results of four published reports (748 arthroplasties) using cementless femoral components of a tapered design were reviewed. The review found a low incidence of aseptic loosening (0.5%) and significant thigh pain (0.5%), and a radiographic incidence of proximal bone atrophy of 6%. In no case was the metaphyseal bone atrophy described as severe or extensive. These findings challenge the premise that reactive bone adaptation is related to the issues of femoral component stiffness exclusively and propose that femoral component geometry may have an influence on bone adaptation.Keywords
This publication has 15 references indexed in Scilit:
- Skeletal Response to Well Fixed Femoral Components Inserted With and Without CementClinical Orthopaedics and Related Research, 1996
- Clinical and radiographic outcome of a cementless, titanium, plasma spray-coated total hip arthroplasty femoral componentThe Journal of Arthroplasty, 1996
- A Tapered Titanium Femoral Stem Inserted without Cement in a Total Hip Arthroplasty. Radiographic Evaluation and Stability*Journal of Bone and Joint Surgery, 1996
- Uncemented total hip arthroplasty using the CLS stem: A titanium alloy implant with a corundum blast finishThe Journal of Arthroplasty, 1996
- The Influence of Clinical Factors on Periprosthetic Bone RemodelingClinical Orthopaedics and Related Research, 1996
- CLS press-fit total hip arthroplastyTechniques in Orthopaedics, 1991
- Porous-coated hip replacement. The factors governing bone ingrowth, stress shielding, and clinical resultsThe Journal of Bone and Joint Surgery. British volume, 1987
- Viscoelastic properties of wet cortical bone—II. Relaxation mechanismsJournal of Biomechanics, 1979
- Viscoelastic properties of wet cortical bone—I. Torsional and biaxial studiesJournal of Biomechanics, 1979
- Traumatic Arthritis of the Hip after Dislocation and Acetabular FracturesJournal of Bone and Joint Surgery, 1969