Parameters in the design of oxygen detecting oxidase enzyme electrodes
- 1 May 1996
- journal article
- research article
- Published by Wiley in Electroanalysis
- Vol. 8 (5) , 407-413
Abstract
The response of an amperometric oxidase enzyme electrode, monitored by the consumption of oxygen, was numerically modeled using a two substrate model. The reduction current was found to be sensitive to the thiele modulus, Φ, which is dependent on the square of the biorecognition matrix thickness and the concentration of active enzyme within this matrix. The detectable concentration range to which the sensor can be used can be extended by reducing the membrane thickness or the enzyme loading, both of which reduce the signal change with substrate additions. The dependence of Φ on the square of the membrane thickness makes the sensor design exceedingly sensitive to minor changes in thickness and hence manufacturing of reproducible sensors with common deposition methods exceedingly difficult.Keywords
This publication has 18 references indexed in Scilit:
- Improved model of a polypyrrole glucose oxidase modified electrodeJournal of Electroanalytical Chemistry, 1995
- Redox enzyme linked electrochemical sensors: Theory meets practiceMicrochimica Acta, 1995
- Model for an Immobilized Oxidase Enzyme Electrode in the Presence of Two OxidantsAnalytical Chemistry, 1994
- Modeling of amperometric biosensors by a finite-volume methodJournal of Electroanalytical Chemistry, 1993
- Theoretical evaluation of mediation efficiency in enzyme-incorporated electrodesAnalytical Chemistry, 1993
- Modelling of processes in enzyme electrodesBiosensors and Bioelectronics, 1993
- Digital simulation of homogeneous enzyme kinetics for amperometric redox-enzyme electrodesAnalytica Chimica Acta, 1992
- Modeling the kinetics of immobilized glucose oxidaseBiotechnology and Bioengineering, 1987
- Model of a two-substrate enzyme electrode for glucoseAnalytical Chemistry, 1984
- Model for the amperometric enzyme electrode obtained through digital simulation and applied to the immobilized glucose oxidase systemAnalytical Chemistry, 1975