Signal correction by detection of scanning position in a white-light interferometer for exact surface profile measurement
- 29 April 2019
- journal article
- research article
- Published by Optica Publishing Group in Applied Optics
- Vol. 58 (13) , 3548-3554
Abstract
In order to perform an exact surface profile measurement with a white-light scanning interferometer (WLSI), an actual optical path difference (OPD) changing with time is detected with an additional interferometer in which the light source of the WLSI and an optical band-pass filter are used. This interferometer is simply equipped in the WLSI and does not negatively influence the WLSI. The real OPD is easily calculated from an interference signal with the same signal processing as that in the WLSI. The interference signal of the WLSI is corrected with the real OPD values or the real scanning position values. The corrected interference signal with a constant sampling interval is obtained with an interpolation method. With this correction method, a surface profile with a step shape of 3-μm height is measured accurately with an error less than 2 nm.Keywords
Funding Information
- National Natural Science Foundation of China (61575070)
This publication has 13 references indexed in Scilit:
- Passive vibration compensation in scanning white-light interferometryApplied Optics, 2016
- Front Matter: Volume 92762014
- Long-working-distance incoherent-light interference microscopeApplied Optics, 2005
- High-stability white-light interferometry with reference signal for real-time correction of scanning errorsOptical Engineering, 2003
- High-precision shape measurement by white-light interferometry with real-time scanner error correctionApplied Optics, 2002
- Determination of fringe order in white-light interference microscopyApplied Optics, 2002
- Self-calibration of a scanning white light interference microscopeOptical Engineering, 2000
- Improved vertical-scanning interferometryApplied Optics, 2000
- White light phase-shifting interferometry with self-compensation of PZT scanning errors1999
- Surface Profiling by Analysis of White-light Interferograms in the Spatial Frequency DomainJournal of Modern Optics, 1995