Abstract
A new incus-body driving type transducer relying on piezoelectric stack, with broad frequency bandwidth, is proposed for use in a middle ear implant. To aid the design process of this transducer, a coupling biomechanical model of the human middle ear and the piezoelectric transducer was established by reverse engineering technology. The validity of this model was confirmed by comparing model predicted motions with experimental measurements. Based on this verified biomechanical model, the main parameters of the transducer were determined. And its power consumption was calculated. Finally, to verify the capability of the designed piezoelectric transducer, a human temporal bone experimental platform was built. And the dynamic characteristics and the stimulated performance of the piezoelectric transducer were tested. The result showed that stapes displacement stimulated by the transducer excitation at 10.5 V RMS was equivalent to that from acoustic stimulation at 100 dB SPL, which is an adequate stimulation to the ossicular chain. The corresponding power consumption is 0.31 mW per volt of excitation at 1 kHz, which is low enough for the transducer to be used in a middle ear implant. Besides, this transducer demonstrates high performance at high frequencies.Keywords
Funding Information
- National Natural Science Foundation of China (51305442, BK20130194, 20130095120010, 2012QNB16)
- Jiangsu Provincial Natural Science Foundation (51305442, BK20130194, 20130095120010, 2012QNB16)
- Doctoral Program of Higher Education of China (51305442, BK20130194, 20130095120010, 2012QNB16)
- Fundamental Research Funds for the Central Universities (51305442, BK20130194, 20130095120010, 2012QNB16)
- Priority Academic Program Development of Jiangsu Higher Education Institutions (51305442, BK20130194, 20130095120010, 2012QNB16)
This publication has 27 references indexed in Scilit:
- A Biomechanical Model of the Inner Ear: Numerical Simulation of the Caloric TestThe Scientific World Journal, 2013
- Finite element analysis of the coupling between ossicular chain and mass loading for evaluation of implantable hearing deviceHearing Research, 2011
- Finite element model of human ear reconstruction through micro-computer tomographyActa Oto-Laryngologica, 2010
- Effects of ear-canal pressurization on middle-ear bone- and air-conduction responsesHearing Research, 2010
- A totally implantable hearing system – Design and function characterization in 3D computational model and temporal bonesHearing Research, 2010
- The EarLens system: New sound transduction methodsHearing Research, 2010
- Middle Ear Implantable Hearing Devices: An OverviewTrends in Amplification, 2009
- Ossicular resonance modes of the human middle ear for bone and air conductionThe Journal of the Acoustical Society of America, 2009
- Vibration Modeling and Design of Piezoelectric Floating Mass Transducer for Implantable Middle Ear Hearing DevicesIEICE Transactions on Fundamentals of Electronics, Communications and Computer Sciences, 2007
- Computer-integrated finite element modeling of human middle earBiomechanics and Modeling in Mechanobiology, 2002