Validation of 15 kGy as a radiation sterilisation dose for bone allografts manufactured at the Queensland Bone Bank: application of the VDmax 15 method
Authors:
- 5 March 2008
- journal article
- research article
- Published by Springer Nature in Cell and Tissue Banking
- Vol. 9 (2) , 139-147
Abstract
Background ISO 11137-2006 (ISO 11137-2a 2006) provides a VDmax 15 method for substantiation of 15 kGy as radiation sterilisation dose (RSD) for health care products with a relatively low sample requirement. Moreover, the method is also valid for products in which the bioburden level is less than or equal to 1.5. In the literature, the bioburden level of processed bone allografts is extremely low. Similarly, the Queensland Bone Bank (QBB) usually recovers no viable organisms from processed bone allografts. Because bone allografts are treated as a type of health care product, the aim of this research was to substantiate 15 kGy as a RSD for frozen bone allografts at the QBB using method VDmax 15—ISO 11137-2: 2006 (ISO 11137-2e, Procedure for method VDmax 15 for multiple production batches. Sterilisation of health care products – radiation – part 2: establishing the sterilisation dose, 2006; ISO 11137-2f, Procedure for method VDmax 15 for a single production batch. Sterilisation of health care products – radiation – part 2: establishing the sterilisation dose, 2006). Materials 30 femoral heads, 40 milled bone allografts and 40 structural bone allografts manufactured according to QBB standard operating procedures were used. Method Estimated bioburdens for each bone allograft group were used to calculate the verification doses. Next, 10 samples per group were irradiated at the verification dose, sterility was tested and the number of positive tests of sterility recorded. If the number of positive samples was no more than 1, from the 10 tests carried out in each group, the verification was accepted and 15 kGy was substantiated as RSD for those bone allografts. Results The bioburdens in all three groups were 0, and therefore the verification doses were 0 kGy. Sterility tests of femoral heads and milled bones were all negative (no contamination), and there was one positive test of sterility in the structural bone allograft. Accordingly, the verification was accepted. Conclusion Using the ISO validated protocol, VDmax 15, 15 kGy was substantiated as RSD for frozen bone allografts manufactured at the QBB.Keywords
This publication has 14 references indexed in Scilit:
- Sterilization of allograft bone: effects of gamma irradiation on allograft biology and biomechanicsCell and Tissue Banking, 2006
- Sterilization of Allograft Bone: is 25 kGy the Gold Standard for Gamma Irradiation?Cell and Tissue Banking, 2006
- Validation and Substantiation of 25 kGy as Sterilization Dose for Lyophilized Human Amnion MembraneCell and Tissue Banking, 2006
- Validating a Low Dose Gamma Irradiation Process for Sterilizing Allografts Using ISO 11137 Method 2BCell and Tissue Banking, 2005
- Hospital-based allogenic bone bank—10-year experienceJournal of Hospital Infection, 2005
- Clostridium Infections Associated with Musculoskeletal-Tissue AllograftsNew England Journal of Medicine, 2004
- Indonesia: Statistical Sampling Technique in Validation of Radiation Sterilisation Dose of Biological TissueCell and Tissue Banking, 2003
- Experience with a bone bank operation and allograft bone infection in recipients at a medical centre in southern TaiwanJournal of Hospital Infection, 2002
- Bacterial contamination of femoral head allografts from living donorsActa Orthopaedica, 2002
- Infection associated with the use of allograft bone from the North East Scotland Bone BankJournal of Hospital Infection, 1997