The Imaging X-ray Polarimetry Explorer (IXPE) is a NASA Small Explorer mission—in partnership with the Italian Space Agency—dedicated to X-ray polarimetry in the 2–8 keV energy band. The IXPE telescope comprises three grazing incidence mirror modules coupled to three detector units hosting each one a Gas Pixel Detector, a gas detector that allows measuring the polarization degree by using the photoelectric effect. A wide and accurate ground calibration was carried out on the IXPE Detector Units at INAF-IAPS, in Italy, where a dedicated facility was setup at this aim. In this paper, we present the results obtained from this calibration campaign to study the IXPE focal plane detector response to polarized radiation. In particular, we report on the modulation factor, which is the main parameter to estimate the sensitivity of a polarimeter.Keywords
All Related Versions
Funding Information
- Agenzia Spaziale Italiana (ASI-OHBI-2017-12-I.0)
- Agenzia Spaziale Italiana (ASI-INAF-2017-12-H0)
- Agenzia Spaziale Italiana (ASI-INFN- 2017.13-H0)
This publication has 27 references indexed in Scilit:
- Performance of the Gas Pixel Detector: an x-ray imaging polarimeter for upcoming missions of astrophysics2016
- The Imaging X-ray Polarimetry Explorer (IXPE)2016
- Analyzing the data from X-ray polarimeters with Stokes parametersAstroparticle Physics, 2015
- ON THE OPERATION OF X-RAY POLARIMETERS WITH A LARGE FIELD OF VIEWThe Astrophysical Journal, 2014
- ON THE STATISTICAL ANALYSIS OF X-RAY POLARIZATION MEASUREMENTSThe Astrophysical Journal, 2013
- A sealed Gas Pixel Detector for X-ray astronomyNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2007
- Direct reading of charge multipliers with a self-triggering CMOS analog chip with 105k pixels at 50μm pitchNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2006
- An efficient photoelectric X-ray polarimeter for the study of black holes and neutron starsNature, 2001
- A precision measurement of the X-ray polarization of the Crab Nebula without pulsar contaminationThe Astrophysical Journal, 1978
- Measurement of the X-ray polarization of the Crab NebulaThe Astrophysical Journal, 1976