Comparison of simultaneous continuous intracranial pressure (ICP) signals from ICP sensors placed within the brain parenchyma and the epidural space
- 2 March 2007
- journal article
- research article
- Published by IOP Publishing in Medical Engineering & Physics
- Vol. 30 (1) , 34-40
Abstract
Simultaneous continuous intracranial pressure (ICP) signals from two different ICP sensors (Codman ICP MicroSensor; Johnson & Johnson, Raynham, MA) placed within the brain parenchyma and the epidural space were compared in ten patients with idiopathic normal pressure hydrocephalus (iNPH). Comparisons were made at the single ICP wave level by determining differences in mean pressure, and pulse pressure amplitude (dP) and latency (dT, i.e. rise time). Differences for the parameters mean ICP, mean ICP wave amplitude and mean ICP wave latency were also determined during consecutive 6-s time windows. The ICP sensors located within the brain parenchyma and epidural space showed marked differences in mean pressure. On the contrary, there were minor differences between sensors in 8 of 10 patients regarding single wave pulse pressure amplitude (dP) and single wave latency (dT, i.e. rise time), and also with regard to the parameters mean ICP wave amplitude and mean ICP wave latency. The levels of mean ICP wave amplitudes used for predicting shunt response in iNPH patients were independent of sensor location in 9 of 10 patients. These results extend previous data that epidural ICP monitoring cannot be used to reliably determine mean ICP, however, epidural ICP monitoring is very useful for determining ICP waveform parameters such as pulse pressure amplitude (dP) and mean ICP wave amplitude. It is suggested that epidural ICP monitoring with determination of mean ICP wave amplitude can be used to predict shunt response in iNPH, though an ICP sensor designed for epidural ICP monitoring would be preferable.Keywords
This publication has 17 references indexed in Scilit:
- Intracranial pressure parameters in idiopathic normal pressure hydrocephalus patients treated with ventriculo-peritoneal shuntsActa Neurochirurgica, 2005
- A new method for processing of continuous intracranial pressure signalsMedical Engineering & Physics, 2005
- A prospective study on the clinical effect of surgical treatment of normal pressure hydrocephalus: the value of hydrodynamic evaluationBritish Journal Of Neurosurgery, 2004
- Reliability of Epidural Pressure Measurement in Clinical Practice: Behavior of Three Modern Sensors during Simultaneous Ipsilateral Intraventricular or Intraparenchymal Pressure MeasurementNeurosurgery, 1998
- A comparison of extradural and intraparenchymatous intracranial pressures in head injured patientsIntensive Care Medicine, 1995
- Comparative analysis between epidural (Gaeltec) and suboural (Camino) intracranial pressure probesJournal of Clinical Monitoring and Computing, 1992
- Clinical evaluation of a simple epidural pressure sensorActa Neurochirurgica, 1986
- Behavior of an extradural pressure monitor in clinical useJournal of Neurosurgery, 1985
- Simultaneous Recording of the Ventricular Fluid Pressure and the Epidural PressureEuropean Neurology, 1972
- MINIATURE TRANSDUCER FOR INTRACRANIAL PRESSURE MONITORING IN MANActa Neurologica Scandinavica, 1970