A Rare Case of Persistent Bacteremia: Leadless Micra Pacemaker Endocarditis
Authors:
Open Access
- 18 January 2023
- journal article
- case report
- Published by Wiley in Case Reports in Cardiology
- Vol. 2023, 1-5
Abstract
Leadless pacing systems have revolutionized the field of electrophysiology given its low complication rates and almost non-existent rate of infections compared with traditional pacemakers. These devices boast resistance to infections given its unique features; however, as described in this report, device-related infection from these leadless devices is still possible. In patients with leadless pacing system that is persistently bacteremic in the future, evaluation of the device with transesophageal echocardiogram or intracardiac echocardiography should be performed, and if vegetation is noted on the device, device extraction should highly be considered, along with empiric intravenous antibiotics. Lastly, new leadless device should not be re-implanted within 2 weeks of the removal of the infected device to prevent seeding of the new device.Keywords
All Related Versions
This publication has 10 references indexed in Scilit:
- Safety and Efficacy of Leadless Pacemakers: A Systematic Review and Meta‐AnalysisJournal of the American Heart Association, 2021
- Retrieval of an infected leadless pacemakerHeartRhythm Case Reports, 2020
- Leadless pacemakers reduce risk of device-related infection: Review of the potential mechanismsHeart Rhythm, 2020
- Reduced bacterial adhesion with parylene coating: Potential implications for Micra transcatheter pacemakersJournal of Cardiovascular Electrophysiology, 2020
- Quality of life of patients undergoing conventional vs leadless pacemaker implantation: A multicenter observational studyJournal of Cardiovascular Electrophysiology, 2019
- Health‐related quality of life impact of a transcatheter pacing systemJournal of Cardiovascular Electrophysiology, 2018
- A leadless pacemaker in the real-world setting: The Micra Transcatheter Pacing System Post-Approval RegistryHeart Rhythm, 2017
- Treating an infected transcatheter pacemaker system via percutaneous extractionHeartRhythm Case Reports, 2016
- Transvenous Lead Extraction: Heart Rhythm Society Expert Consensus on Facilities, Training, Indications, and Patient ManagementHeart Rhythm, 2009