Relationship between file size and stiffness of stainless steel instruments
Authors:
- 1 December 1994
- journal article
- research article
- Published by Wiley in Dental Traumatology
- Vol. 10 (6) , 260-263
Abstract
The purpose of this study was to evaluate the relationship existing between file size and stiffness for three endodontic files of different cross-sections. Three groups of instruments were tested: Flexofiles from the Maillefer company, K Files and K-Flex from the Kerr company. Flexofiles have a triangular cross-section, K Files a square cross-section, and K-Flex instruments a rhombus cross-section from size 15 to size 40. For each group, 10 files from size 15 to size 40 were compared according to ANSI/ADA specification N degrees 28 for bending moment evaluation. There was a statistical difference between the three groups: the square cross-section K files presented a larger bending moment than the rhombus cross-section K Flex, which presented a higher bending moment than the triangular cross-section Flexofiles. There was an exponential relationship between file size and bending moment.Keywords
This publication has 16 references indexed in Scilit:
- Torsional and stiffness properties of canal master U stainless steel and nitinol instrumentsJournal of Endodontics, 1994
- Another look at root canal instrumentationOral Surgery, Oral Medicine, Oral Pathology, 1992
- The “balanced force” concept for instrumentation of curved canalsJournal of Endodontics, 1985
- Bending and torsion of endodontic files with rhombus cross sectionsJournal of Endodontics, 1982
- Flute design of endodontic instruments: its influence on cutting efficiencyJournal of Endodontics, 1982
- The Anticurvature Filing Method to Prepare the Curved Root CanalThe Journal of the American Dental Association, 1980
- The influence of the method of canal preparation on the quality of apical and coronal obturationJournal of Endodontics, 1979
- Brittle and ductile torsional failures of endodontic instrumentsJournal of Endodontics, 1977
- Cutting efficiency of endodontic reamersOral Surgery, Oral Medicine, Oral Pathology, 1973
- Bending and torsion properties of endodontic instrumentsOral Surgery, Oral Medicine, Oral Pathology, 1968