A New Constant Deviation Monochromator for Angle Resolved Photoemission Spectroscopy

Authors:
  • TMTsuneaki Miyahara
    Tsuneaki Miyahara
  • SSShoji Suzuki
    Shoji Suzuki
  • THTakaaki Hanyu
    Takaaki Hanyu
  • HKHiroo Kato
    Hiroo Kato
  • KNKenzo Naito
    Kenzo Naito
  • HFHirohito Fukutani
    Hirohito Fukutani
  • INIchiro Nagakura
    Ichiro Nagakura
  • HSHidenao Sugawara
    Hidenao Sugawara
  • SNShun-Ichi Nakai
    Shun-Ichi Nakai
  • TITakehiko Ishii
    Takehiko Ishii
  • HNHideyuki Noda
    Hideyuki Noda
  • TNTakeshi Namioka
    Takeshi Namioka

Abstract
A new type of constant deviation monochromator equipped with three interchangeable concave gratings (600, 1200 and 2400 grooves/mm) of 2 m radius of curvature was designed and built specifically for angle resolved photoemission spectroscopy at photon energies ranging from 20 to 160 eV. The exit slit of the monochromator is at a fixed position, whereas the entrance slit moves, simultaneously with grating rotation, along the direction of the incident light so as to minimize the amount of defocusing. A monochromatic light spot of 0.5 mm in diamter was obtained at the fixed sample position with the aid of a toroidal postfocussing mirror. It is found over the energy range concerned that the resolution is bettern than 1700 with 25 µm slits, and the photon flux obtained behind the toroidal mirror for a horizontal acceptance angle 0.5 mrad is 4×1010 photons/s when the width of the slits is 50 µm and the stored current for synchrotron radiation is 100 mA.