Pre‐excitation gradients for eddy current nulled convex optimized diffusion encoding (Pre‐ENCODE)
- 19 March 2024
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 92 (2) , 573-585
Abstract
Purpose: To evaluate the use of pre‐excitation gradients for eddy current–nulled convex optimized diffusion encoding (Pre‐ENCODE) to mitigate eddy current–induced image distortions in diffusion‐weighted MRI (DWI).Methods: DWI sequences using monopolar (MONO), ENCODE, and Pre‐ENCODE were evaluated in terms of the minimum achievable echo time (TE) and eddy current‐induced image distortions using simulations, phantom experiments, and in vivo DWI in volunteers ().Results: Pre‐ENCODE provided a shorter TE than MONO (71.0 17.7ms vs. 77.6 22.9ms) and ENCODE (71.0 17.7ms vs. 86.2 14.2ms) in 100 of the simulated cases for a commercial 3T MRI system with b‐values ranging from 500 to 3000 s/mm and in‐plane spatial resolutions ranging from 1.0 to 3.0mm. Image distortion was estimated by intravoxel signal variance between diffusion encoding directions near the phantom edges and was significantly lower with Pre‐ENCODE than with MONO (10.1 vs. 22.7, ) and comparable to ENCODE (10.1 vs. 10.4, ). In vivo measurements of apparent diffusion coefficients were similar in global brain pixels (0.37 [0.28,1.45] mm/s vs. 0.38 [0.28,1.45]mm/s, ) and increased in edge brain pixels (0.80 [0.17,1.49] mm/s vs. 0.70 [0.18,1.48]mm/s, ) for MONO compared to Pre‐ENCODE.Conclusion: Pre‐ENCODE mitigated eddy current–induced image distortions for diffusion imaging with a shorter TE than MONO and ENCODE.Keywords
Funding Information
- National Institute of Mental Health (R01‐MH116173)
- National Institute of Biomedical Imaging and Bioengineering (U01‐EB025162)
This publication has 35 references indexed in Scilit:
- Longitudinal gradient coil optimization in the presence of transient eddy currentsMagnetic Resonance in Medicine, 2007
- Comprehensive approach for correction of motion and distortion in diffusion‐weighted MRIMagnetic Resonance in Medicine, 2003
- Reduction of eddy‐current‐induced distortion in diffusion MRI using a twice‐refocused spin echoMagnetic Resonance in Medicine, 2002
- Design of improved spectral-spatial pulses for routine clinical useMagnetic Resonance in Medicine, 2000
- Characterization of and correction for eddy current artifacts in echo planar diffusion imagingMagnetic Resonance in Medicine, 1998
- Concomitant gradient terms in phase contrast MR: Analysis and correctionMagnetic Resonance in Medicine, 1998
- Elimination of eddy current artifacts in diffusion‐weighted echo‐planar images: The use of bipolar gradientsMagnetic Resonance in Medicine, 1997
- Consistent fat suppression with compensated spectral‐spatial pulsesMagnetic Resonance in Medicine, 1997
- Correction for distortion of echo‐planar images used to calculate the apparent diffusion coefficientMagnetic Resonance in Medicine, 1996
- Temporal and spatial analysis of fields generated by eddy currents in superconducting magnets: Optimization of corrections and quantitative characterization of magnet/gradient systemsMagnetic Resonance in Medicine, 1991