Wear Morphology of Metal-Metal Implants: Hip Simulator Tests Compared with Clinical Retrievals
Authors:
- 1 January 1998
- book chapter
- Published by ASTM International
- p. 129-143
Abstract
This study compared the bearing surfaces of modern-generation metal- metal implants worn in five different hip joint simulators and worn in vivo. Comparable wear morphology was found between in vitro and in vivo worn implants. The non-contact zones displayed fine residual scratches from the original polishing, and carbide bumps that were either rounded or irregular. In the main bearing zones, the polishing marks and third- body scratches were worn smooth. Moreover, there were flat-bottomed depressions, apparently carbides that had been dissolved below the level of surrounding matrix. Localized numerous micropits were found in the smoothly worn area on most of simulator tested specimens and on some of the retrieved implants. Surrounding the main wear zone was a transition zone containing dense third-body scratches, and thin deposits containing calcium phosphate. The comparable wear morphology suggested that in vitro wear simulators can provide an accurate model of the wear in vivo of modern-generation metal-metal total hip replacements.Keywords
This publication has 5 references indexed in Scilit:
- Clinical Wear Performance of Metal-on-Metal Hip Arthroplasties1998
- Frictional heating of bearing materials tested in a hip joint wear simulatorProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1997
- In Vivo Wear of 3 Types of Metal on Metal Hip Prostheses During 2 Decades of UseClinical Orthopaedics and Related Research, 1996
- The McKee-Farrar hip arthroplasty. A long-term studyThe Journal of Bone and Joint Surgery. British volume, 1986
- The Wear on Removed McKee-Farrar Total Hip ProsthesesJournal of Bone and Joint Surgery, 1974