Stability and accuracy of active shielding for grounded capacitive sensors
Authors:
- 28 September 2006
- journal article
- Published by IOP Publishing in Measurement Science and Technology
- Vol. 17 (11) , 2884-2890
Abstract
Active shielding is commonly used to measure remote grounded capacitive sensors because it reduces the effects of both external noise/interference and parasitic capacitances of the shielded cable. However, due to active shielding, the measurement circuit can become unstable and inaccurate. This paper analyses these limitations theoretically and experimentally, and then provides guidelines for improving the performance of active shielding. One of the key points is the selection of the bandwidth of the amplifier that drives the shield of the coaxial cable. A wide bandwidth improves accuracy, but a narrow bandwidth improves stability. Therefore, there is a trade-off between stability and accuracy with respect to the bandwidth of the amplifier.Keywords
This publication has 5 references indexed in Scilit:
- A Low-Cost Noncontact Capacitance-Type Level Transducer for a Conducting LiquidIEEE Transactions on Instrumentation and Measurement, 2006
- Integrated Interfaces for Low-Cost Multiple-Sensor SystemsJournal of Intelligent Material Systems and Structures, 1999
- High-accuracy measurement techniques for capacitance transducersMeasurement Science and Technology, 1993
- Electronic transducers for industrial measurement of low value capacitancesJournal of Physics E: Scientific Instruments, 1988
- A water-level sensor using a capacitance to frequency converterJournal of Physics E: Scientific Instruments, 1983