Computing patient-specific hemodynamics in stented femoral artery models obtained from computed tomography using a validated 3D reconstruction method
- 1 November 2019
- journal article
- research article
- Published by IOP Publishing in Medical Engineering & Physics
- Vol. 75 (1) , 23-35
Abstract
Patients with peripheral artery disease who undergo endovascular treatment are often inflicted by in-stent restenosis. The relation between restenosis and abnormal hemodynamics may be analyzed using patient-specific computational fluid dynamics (CFD) simulations. In this work, first a three-dimensional (3D) reconstruction method, based on an in-house semi-automatic segmentation algorithm of a patient's computed tomography (CT) images with calcification and metallic artifacts, and thrombus removal is described. The reconstruction method was validated using 3D printed rigid phantoms of stented femoral arteries by comparing the reconstructed geometries with the reference computer-aided design (CAD) geometries employed for 3D printing. The mean reconstruction error resulting from the validation of the reconstruction method was ~6% in both stented and non-stented regions. Secondly, a patient-specific model of the stented femoral artery was created and CFD analyses were performed with emphasis on the selection of the boundary conditions. CFD results were compared in scenarios with and without common femoral artery bifurcation, employing flat or parabolic inlet velocity profiles. Similar helical flow structures were visible in all scenarios. Negligible differences in wall shear stress (<0.5%) were found in the stented region. In conclusion, a robust method, applicable to patient-specific cases of stented diseased femoral arteries, was developed and validated.Keywords
Funding Information
- Fondazione Cariplo (2017-0792)
This publication has 44 references indexed in Scilit:
- CT Imaging of Coronary Stents: Past, Present, and FutureISRN Cardiology, 2012
- Hemodynamics and In-stent Restenosis: Micro-CT Images, Histology, and Computer SimulationsAnnals of Biomedical Engineering, 2011
- Dual-Source CT Angiography of Peripheral Arterial Stents: In Vitro Evaluation of 22 Different Stent TypesRadiology Research and Practice, 2011
- Wall Shear Stress in the Superficial Femoral Artery of Healthy Adults and its Response to Postural Changes and ExerciseEuropean Journal of Vascular and Endovascular Surgery, 2011
- Helical flow as fluid dynamic signature for atherogenesis risk in aortocoronary bypass. A numeric studyJournal of Biomechanics, 2007
- Curvature and tortuosity of the superficial femoral artery: a possible risk factor for peripheral arterial diseaseJournal of Applied Physiology, 2006
- In stent restenosis: bane of the stent eraJournal of Clinical Pathology, 2006
- The future of ultrasound: viewing the dark side of the moon?Ultrasound in Obstetrics & Gynecology, 2004
- Magnetic resonance imaging measurement of blood volume flow in peripheral arteries in healthy subjectsJournal of Vascular Surgery, 2003
- Active contours without edgesIEEE Transactions on Image Processing, 2001