Reducing gradient imperfections for spiral magnetic resonance spectroscopic imaging
Authors:
Open Access
- 24 May 2006
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 56 (1) , 198-203
Abstract
Spiral k-space magnetic resonance spectroscopic imaging (MRSI) requires high performance from gradient hardware systems. During the readout phase, oscillating gradients are continuously played out, which can cause undesired effects. These effects on the quality of SI data are non-intuitive because of their time-varying nature. In this work we describe the effects of undesirable gradient performance on SI. Measurements of the true readout trajectories were performed and the results were then used in the reconstruction process. The effects of these imperfections resulted in a spatially and spectrally varying amplitude and frequency modulation. The use of the measured trajectories in the reconstruction process yielded an up to 20% increase in signal amplitude recovery. Magn Reson Med, 2006.Keywords
This publication has 15 references indexed in Scilit:
- Detection and correction of frequency instabilities for volumetric 1H echo‐planar spectroscopic imagingMagnetic Resonance in Medicine, 2005
- Spiral readout gradients for the reduction of motion artifacts in chemical shift imagingMagnetic Resonance in Medicine, 2004
- Phase coherent averaging in magnetic resonance spectroscopy using interleaved navigator scans: Compensation of motion artifacts and magnetic field instabilitiesMagnetic Resonance in Medicine, 2002
- Volumetric spectroscopic imaging with spiral‐based k‐space trajectoriesMagnetic Resonance in Medicine, 1998
- A Method to measure arbitrary k‐space trajectories for rapid MR imagingMagnetic Resonance in Medicine, 1997
- Compensation of multi‐dimensional selective excitation pulses using measured k‐space trajectoriesMagnetic Resonance in Medicine, 1995
- Quantitative characterization of the eddy current fields in a 40‐cm bore superconducting magnetMagnetic Resonance in Medicine, 1994
- Analysis of the eddy-current induced artifacts and the temporal compensation in nuclear magnetic resonance imagingIEEE Transactions on Medical Imaging, 1991
- In vivo proton spectroscopy in presence of eddy currentsMagnetic Resonance in Medicine, 1990
- A method of measuring field-gradient modulation shapes. Application to high-speed NMR spectroscopic imagingJournal of Physics E: Scientific Instruments, 1987