Polyethylene Liner Dissociation Is a Complication of the DePuy Pinnacle Cup: A Report of 23 Cases
- 1 February 2016
- journal article
- research article
- Published by Wolters Kluwer Health in Clinical Orthopaedics and Related Research
- Vol. 474 (2) , 441-446
Abstract
Polyethylene liner dissociation is a rare but catastrophic event in total hip arthroplasty (THA), and certain implant designs are known to be at greater risk. Although the DePuy Pinnacle (Warsaw, IN, USA) modular acetabular construct has an excellent record of fixation and wear, an unexpectedly high number of liner dissociations has been noted. The purposes of this study were (1) to characterize the clinical parameters observed in a large group of patients who have experienced liner dissociations with the DePuy Pinnacle acetabular component; (2) to describe the radiographic findings in this group of patients; and (3) to calculate a minimum frequency of this complication. Since 2001, 23 patients with previously well-functioning THAs presented with sudden atraumatic polyethylene liner dissociation at four separate institutions. These THAs were performed between 2001 and 2013. Eight different arthroplasty specialists had performed the index hip arthroplasties using the DePuy Pinnacle acetabular component with a polyethylene liner. Polyethylene failures were evaluated for liner type and radiographic cup position. For three of the surgeons who contributed cases, institutional registries allowed the calculation of the number of components of this type that they used during the period in question, which provided a conservative estimate of the frequency of this type of failure. All 23 liner failures occurred atraumatically in previously asymptomatic THAs at a mean of 48 months (range, 3–138 months). Patients characteristically reported a new and sudden onset of discomfort with audible, reproducible squeaking. Surgical inspection of dissociated liners demonstrated displacement of polyethylene with shearing of the peripheral locking tabs. Radiographic evaluation demonstrated that 14 cups were well positioned and nine cups were malpositioned outside the so-called safe zone. Conservative estimates of the frequency of this complication from the three surgeons’ practices whose institutional registries allowed calculation of the lowest possible frequency were 0.32% (six of 1888), 0.77% (three of 391), and 0.82% (three of 367). With this report of 23 additional liner dissociations, we suggest that surgeons should be aware of the problem and take extra precautions when using this implant to ensure locking mechanism integrity at the time of surgery. We caution that the frequency of liner dissociation may be higher than previously reported. Level IV, therapeutic study.Keywords
This publication has 17 references indexed in Scilit:
- Independent predictors of failure up to 7.5 years after 35 386 single-brand cementless total hip replacementsThe Bone & Joint Journal, 2013
- Spontaneous Dissociation of Offset, Face-Changing Polyethylene Liners from the Acetabular ShellJournal of Bone and Joint Surgery, 2012
- Late Liner Disassociation of a Pinnacle System Acetabular ComponentOrthopedics, 2012
- Five Duraloc Locking Ring FailuresThe Journal of Arthroplasty, 2010
- A Comparison of a Second- and a Third-Generation Modular Cup Design: Is New Improved?The Journal of Arthroplasty, 2010
- Acute Liner Disassociation of a Pinnacle Acetabular ComponentThe Journal of Arthroplasty, 2009
- Impingement as a Mechanism of Dissociation of a Metasul Metal-on-Metal LinerThe Journal of Arthroplasty, 2009
- Analysis and Results of Dissociation of the Polyethylene Liner in the Harris-Galante I Acetabular ComponentThe Journal of Arthroplasty, 2008
- A Review of Current Cross-linked Polyethylenes Used in Total Joint ArthroplastyClinical Orthopaedics and Related Research, 2005
- Late disassembly of modular acetabular components: A report of two casesActa Orthopaedica, 1993