Experimental validation of noninvasive referencing in navigated procedures on long bones
- 6 November 2006
- journal article
- research article
- Published by Wiley in Journal of Orthopaedic Research
- Vol. 25 (2) , 201-207
Abstract
Navigation procedures in orthopedic surgery require fixation of reference markers to the anatomic region of interest. Inadequate fixation might lead to micromotion or loosening of the reference marker, consequently causing registration failures or errors in navigation. Osseous rigid fixation is usually achieved by minimally invasive Schanz screws or pins. The goal of this study was to evaluate a non invasive external fixation device, a headband so far used in cranial navigation, as an alternative invasive fixation technique to reference markers in the femur. A common navigation system with an adapted trauma software application was used to track the positions of the soft tissue‐attached headband relative to an invasive reference marker on the femur during manipulations of the thigh. Relative translative and rotational movements of the headband were measured during defined movements of the hip and knee and manipulations of the headband itself. The results revealed high translative and rotational movements, up to 6 mm and 3°, respectively, due to minor manipulations of the affected lower extremity. Noninvasive soft tissue fixation with a headband does not allow rigid fixation for accurate navigated registration or operative procedures at the femur. Necessary intraoperative movements or manipulations would cause substantial registration failures. Invasive fixation techniques with screws or pins are still the method of choice. © 2006 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 25:201–207, 2007Keywords
This publication has 23 references indexed in Scilit:
- epiDRB - ein neues minimal-invasives Konzept für die Referenzierung im Rahmen der Computer-assistierten orthopädischen ChirurgieZeitschrift für Orthopädie und ihre Grenzgebiete, 2005
- Advanced surface-recording techniques for computer-assisted oral and maxillofacial surgeryBritish Journal of Oral and Maxillofacial Surgery, 2004
- Joint replacement—total hip replacement with CT-based navigationInjury, 2004
- Computer aided high tibial open wedge osteotomyInjury, 2004
- Which navigation when?Injury, 2004
- Modality-based navigationInjury, 2004
- EditorialInjury, 2004
- Navigation-Supported Surgery in the Head and Neck RegionLaryngo-Rhino-Otologie, 2003
- Comparative Tracking Error Analysis of Five Different Optical Tracking SystemsComputer Aided Surgery, 2000
- Comparative tracking error analysis of five different optical tracking systemsComputer Aided Surgery, 2000