Oxygen-15 labeled CO2, O2, and CO PET in small animals: evaluation using a 3D-mode microPET scanner and impact of reconstruction algorithms
Open Access
- 27 October 2017
- journal article
- research article
- Published by Springer Nature in EJNMMI Research
- Vol. 7 (1) , 1-9
Abstract
Positron emission tomography (PET) studies using 15O-labeled CO2, O2, and CO have been used in humans to evaluate cerebral blood flow (CBF), the cerebral oxygen extraction fraction (OEF), and the cerebral metabolic rate of oxygen (CMRO2) and cerebral blood volume (CBV), respectively. In preclinical studies, however, PET studies using 15O-labeled gases are not widely performed because of the technical difficulties associated with handling labeled gases with a short half-life. The aims of the present study were to evaluate the scatter fraction using 3D-mode micro-PET for 15O-labeled gas studies and the influence of reconstruction algorithms on quantitative values. Nine male SD rats were studied using the steady state inhalation method for 15O-labeled gases with arterial blood sampling. The resulting PET images were reconstructed using filtered back projection (FBP), ordered-subset expectation maximization (OSEM) 2D, or OSEM 3D followed by maximum a posteriori (OSEM3D-MAP). The quantitative values for each brain region and each reconstruction method were calculated by applying different reconstruction methods. The quantitative values for the whole brain as calculated using FBP were 46.6 ± 12.5 mL/100 mL/min (CBF), 63.7 ± 7.2% (OEF), 5.72 ± 0.34 mL/100 mL/min (CMRO2), and 5.66 ± 0.34 mL/100 mL (CBV), respectively. The CBF and CMRO2 values were significantly higher when the OSEM2D and OSEM3D-MAP reconstruction methods were used, compared with FBP, whereas the OEF values were significantly lower when reconstructed using OSEM3D-MAP. We evaluated the difference in quantitative values among the reconstruction algorithms using 3D-mode micro-PET. The iterative reconstruction method resulted in significantly higher quantitative values for CBF and CMRO2, compared with the values calculated using the FBP reconstruction method.Keywords
Funding Information
- KAKENHI Grant-in-Aid for Scientific Research (S) (24229008)
- KAKENHI Grant-in-Aid for Scientific Research (B) (25861095)
This publication has 32 references indexed in Scilit:
- Quantitative accuracy of MAP reconstruction for dynamic PET imaging in small animalsMedical Physics, 2012
- NEMA NU-04-based performance characteristics of the LabPET-8™ small animal PET scannerPhysics in Medicine & Biology, 2011
- Cerebral Oxygen Metabolism of Rats using Injectable 15O-Oxygen with a Steady-State MethodJournal of Cerebral Blood Flow & Metabolism, 2011
- The effect of activity outside the field-of-view on image signal-to-noise ratio for 3D PET with15OPhysics in Medicine & Biology, 2011
- Cerebellum as the Normal Reference for the Detection of Increased Cerebral Oxygen ExtractionJournal of Cerebral Blood Flow & Metabolism, 2010
- Image-Quality Assessment for Several Positron Emitters Using the NEMA NU 4-2008 Standards in the Siemens Inveon Small-Animal PET ScannerJournal of Nuclear Medicine, 2010
- NEMA NU4-2008 Image Quality Performance Report for the microPET Focus 120 and for Various Transmission and Reconstruction MethodsJournal of Nuclear Medicine, 2009
- Performance Evaluation of the Inveon Dedicated PET Preclinical Tomograph Based on the NEMA NU-4 StandardsJournal of Nuclear Medicine, 2009
- Quantification of Cerebral Blood Flow and Oxygen Metabolism with 3-Dimensional PET and 15O: Validation by Comparison with 2-Dimensional PETJournal of Nuclear Medicine, 2007
- Development of Injectable O-15 Oxygen and Estimation of Rat OEFJournal of Cerebral Blood Flow & Metabolism, 2003